Quality Audit Scope for Medical Device Manufacturing Sites in Tanzania
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Tanzania applies a separate gate for manufacturing sites — the Quality Audit. TMDA’s Guidelines on Submission of Documentation for Quality Audit of Medical Devices and Diagnostics Manufacturing Sites define the scope of application as follows:
§7.1.1: the quality audit applies to sites manufacturing Class C and Class D medical devices and diagnostics, as well as sites manufacturing Class A and Class B invasive and/or sterile products. Audits of Class A and Class B (non-invasive, non-sterile) sites are determined at the time of the marketing authorization review, depending on product risk.
The audit is conducted against the good manufacturing practice (GMP) requirements contained in the ISO 13485 quality management system standard. Once the site passes, a quality audit certificate valid for 3 years is issued (§9.4), and it may be renewed in accordance with the prescribed procedure. The certificate is site-based, not product-based, which means its lifecycle must be managed separately from the product registration itself.
2. Legal basis
| Instrument | Reference / number | Publication date | Provisions |
|---|---|---|---|
| Guidelines on Submission of Documentation for Quality Audit of Medical Devices and Diagnostics Manufacturing Sites | TMDA/DMD//MDL/F/010 | No publication date stated (current TMDA version) | §7.1.1, §7.1.2, §7.2, §8, §9.1–§9.4 |
| Compendium of Guidelines for Marketing Authorization of Medical Devices, Diagnostics and Laboratory Equipment | No Government Gazette number (TMDA guideline document) | June 2024 | Part I §3.4.5 (quality audit integrated into the evaluation process; timelines exclude the on-site inspection) |
| The Tanzania Food, Drugs and Cosmetics (Control of Medical Devices) Regulations, 2015 | GN. No. 315 | 31 July 2015 | Regulation 18(f) (inadequate QMS and GMP as a ground for suspension or cancellation) |
3. Expert insights
Insight 1: Class A/B sites are not exempt across the board — the trigger is the two words invasive and/or sterile.
The wording of §7.1.1 is “class A, B invasive and/or sterile products”. Note the and/or: invasiveness alone triggers the audit, and sterility alone triggers it as well. A sterile syringe (Class B) and an invasive catheter (Class B) both go through the quality audit, whereas an ordinary tongue depressor (Class A) generally does not. GRZAN still performs a dedicated trigger assessment on every Class A/B project, because the classification outcome alone does not answer the question.
Insight 2: The audit scope, the number and qualifications of auditors, and the audit duration are all determined by risk grading.
§7.1.2: TMDA evaluates the application and the accompanying documents and performs a risk assessment; the preliminary evaluation findings and the risk category are used to determine the audit scope, the number and qualifications of auditors, and the number of audit days required. Based on perceived risk, TMDA may also determine whether the site requires an on-site audit or may proceed by desk review. This means that the audit intensity for the same company can differ across product lines, and the outcome cannot be assumed from a previous audit of another product family.
Insight 3: Non-conformities are graded into three levels, and the handling differs completely.
§8: non-conformities are classified as critical, major or minor. The handling rules in §9 are:
- §9.1 Where a critical non-conformity exists, the site must take corrective action and re-apply for an audit as prescribed (re-audit);
- §9.2 Where a major non-conformity exists, the site must take corrective action and submit a compliance report within the timeline set by TMDA (including corrections already implemented and corrective actions to be taken);
- §9.3 For sites assessed through desk review, where a critical or major non-conformity is suspected, the process converts to an on-site audit; where the non-conformity is major or minor, the site can only be found compliant after responding to the query;
- §9.4 Once the requirements are met, the certificate is issued with a validity of 3 years.
Insight 4: Inadequate QMS and GMP is itself a ground for losing the certificate, regardless of whether the product itself is compliant.
Regulation 18(f) of GN 315: based on information obtained after registration, if the QMS and GMP on which the design of Class C and Class D devices is based, or the QMS and GMP on which the manufacture, assembly, processing, packaging, refurbishment or modification of Class B, C and D devices is based, are insufficient to ensure that the device conforms to its specification, TMDA may suspend or cancel the registration. This elevates a manufacturing system deficiency directly into a product licensing problem.
Insight 5: The quality audit certificate is valid for 3 years and is not synchronized with the 5-year registration.
The 3-year validity of the quality audit certificate under §9.4 and the 5-year registration validity under regulation 17(1) of GN 315 are two independent cycles. A company needs to renew the quality audit certificate in year 3 and renew the product registration in year 5 — the two lines must be scheduled separately, with the audit renewal falling roughly in the middle of the registration term.
4. Practical recommendations
- Start quality audit preparation in parallel with any Class C/D project; do not wait until the dossier review is halfway done.
- For Class A/B, perform the invasive/sterile trigger assessment first and record the outcome in the project memorandum.
- Aim for desk review (see article 13 for details), but prepare an on-site audit plan B at the same time — §9.3 provides that desk review converts to on-site audit as soon as a critical or major non-conformity is suspected.
- Maintain two independent calendars — one for the quality audit certificate (3 years) and one for the product registration (5 years) — to avoid missed renewals.
- Ensure the ISO 13485 certificate is issued by a recognized conformity assessment body (CAB) (the parallel requirement for notification applications in Part IV §3.2.3 can be used as a reference).
5. Verification checklist
- □ Product class and invasive/sterile characteristics assessed, and whether the quality audit is triggered confirmed.
- □ Manufacturing site holds a current ISO 13485 certificate issued by a recognized CAB.
- □ Desk review versus on-site audit evaluated, with both options prepared.
- □ Consequences of the three non-conformity grades (critical/major/minor) understood.
- □ Dual calendar established for the quality audit certificate (3 years) and the product registration (5 years).
Regulatory information verified as of 29 October 2026. All provisions cited in this document were checked against the official published texts of TMDA (the quality audit guidelines §7.1.1, §7.1.2, §8, §9.1–§9.4; the June 2024 Compendium of Guidelines, Part I §3.4.5; GN. No. 315, regulation 18(f)), and will be re-checked when the regulations are updated.
- Tanzania local resources: licensed Local Responsible Person (LRP) and distributor resources can be coordinated to cover online submission, sample delivery, quality audit cooperation and port clearance.
- Service capabilities: classification determination, registration and notification pathway selection, technical documentation (CSDT/STED) preparation and localization, GMDN and grouping strategy, clinical evidence organization, software and AI device submissions, change management, vigilance and PMS reporting, annual retention fees and renewal management.
- Contact: hito.lin@grzan.cn; website www.grzan.cn.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us