PPB Guideline HPT/PER/HTR/GUD/011 Rev.02: Scope, Effect and Fees
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
Content verification date: October 30, 2026
1. Key takeaways
- The core operational document for medical device registration in Kenya is the PPB Guidelines for Registration of Medical Devices including In-Vitro Diagnostics, numbered HPT/PER/HTR/GUD/011, version Rev.02, issued in January 2022, effective January 31, 2022, with a review date of January 1, 2027.
- At the end of Section 1.1 the guideline sets its own effect clause: where the guideline conflicts with any written law, the written law prevails. This provision determines the order of argument when a company faces inconsistency between the guideline and the law.
- The scope of application (Section 1.3) has three boundaries: it applies to medical devices and their accessories; devices used to administer medicinal products remain governed by this guideline without prejudice to the medicinal product registration rules; and where a device contains, as an integral component, a substance that on its own may be considered a medicinal product and that acts on the human body in an ancillary manner, the device is to be evaluated and approved under this guideline.
- The guideline does not stand alone. It is accompanied by the Guidelines for Registration of Medical Devices Establishments (PRA/LPP/GUD/072), the Fast Track Review guideline (HPT/PER/GUD/041) and the Suspension and Revocation guideline (HPT/PER/GUD/042).
- The version history is clear: Rev.000 (PPB/PER/MDV/GDL/003, August 2017) → Rev.001 (PPB/PER/MDV/GDL/005, August 2021) → Rev.002 (HPT/PER/HTR/GUD/011, January 2022).
2. Legal basis
| Instrument | Reference / number | Date of publication / entry into force | Provisions / sections | Key points relevant to this article |
|---|---|---|---|---|
| Guidelines for Registration of Medical Devices including In-Vitro Diagnostics | HPT/PER/HTR/GUD/011 Rev.02 | Effective January 31, 2022; review January 1, 2027; signed January 20–25, 2022 | Section 1.1 | Effect clause: written law prevails on conflict; reminder that other applicable legislation must also be observed |
| Same instrument | HPT/PER/HTR/GUD/011 Rev.02 | Same as above | Section 1.2 | Legal framework: Sections 3A(a), 3A, 3B(2)(b), 3B(2)(i), 3B(2)(r) |
| Same instrument | HPT/PER/HTR/GUD/011 Rev.02 | Same as above | Section 1.3 | Three scope points: devices and accessories; drug-administering devices; drug-device combinations (ancillary action) |
| Same instrument | HPT/PER/HTR/GUD/011 Rev.02 | Same as above | Section 4.0 (revision history) | Version history: numbers and dates of Rev.000 / Rev.001 / Rev.002 |
| Guidelines for Registration of Medical Devices Establishments | PRA/LPP/GUD/072 | Published by PPB (date not stated in the text — practical estimate) | Full text | Establishment registration requirements and site inspections |
| Guideline for Fast Track Review of Application for Health Products | HPT/PER/GUD/041 | Published by PPB (date not stated in the text — practical estimate) | Full text | Applicability of fast-track review for Class B and Class C/D products |
| Guideline for Suspension, Withdrawal, Withholding and Revocation | HPT/PER/GUD/042 | Published by PPB (date not stated in the text — practical estimate) | Full text | Circumstances and procedures under which certificates may be suspended, withdrawn, withheld or revoked |
| Pharmacy and Poisons Act | Cap. 244 | Enacted 1957 | Section 3A(a) | Source of the rulemaking power behind the guideline |
3. Expert insights
Insight 1: The written-law-prevails clause is a defence tool in the hands of companies
Section 1.1 of the guideline is blunt: where there is a contradiction between the guideline document and any written law, the latter shall prevail. GRZAN’s reading: this is a favourable clause for companies. The PPB guideline contains extensive operational positions (document lists, route conditions, fee tables), some of which do not appear in Cap. 244 or LN 100 of 2022. When a reviewer raises requirements beyond the written law on the basis of the guideline, companies may cite this clause and request written explanation. In practice, the more robust approach is still to prepare to the guideline standard and argue on the legal standard — leave no gap at the preparation stage, and keep room for reasoned argument if a dispute arises.
Insight 2: Real gaps exist between the guideline and the law; both tracks must be recorded
Fees are the clearest example. The Second Schedule of LN 100 of 2022 lists only four USD tiers for device registration and renewal (USD 100 / 200 / 1,000 / 1,000), while the guideline’s Table 6 additionally distinguishes imported and locally manufactured products and adds a variation fee of USD 200 and annual retention fees (A&B USD 50, C&D USD 300). The two do not contradict each other, but their coverage differs. GRZAN recommends keeping both tables on file for budgeting and submission, with the invoice actually generated by the PRIMS system as the final authority.
Insight 3: Sections 1.3(b) and (c) pull drug-device combinations back onto the device track
Section 1.3(b) provides that devices used to administer medicinal products are governed by this guideline without prejudice to the corresponding registration rules for human and veterinary medicinal products. Section 1.3(c) provides that where a device contains, as an integral component, a substance that on its own may be considered a medicinal product and that acts on the human body in an ancillary manner, the device must be evaluated and approved under this guideline. GRZAN’s reading: Kenya adopts a device-track-first approach with drug rules running in parallel, rather than pushing all combination products to the drug department. This anticipates the Primary Mode of Action (PMOA) determination developed in article 06 — first fix the track, then fix the reviewing department.
Insight 4: The guideline itself is awaiting review; companies should anticipate the revision window
The guideline states a review date of January 1, 2027, and PPB is working toward WHO maturity level 3 and preparing for the WHO global benchmarking tool assessment in September 2026 (public statements — practical estimate). GRZAN’s judgment: the probability of revision around the review milestone is high, and the direction will likely tighten technical documentation requirements and align further with IMDRF. CSDT prepared today should target a higher completeness tier than the current guideline demands, to reduce rework after revision.
4. Practical recommendations
- Write the effect clause into internal appeal templates. Where review requirements exceed the written law, cite Section 1.1 and request written explanation, while still supplying what the guideline asks for.
- File both fee tables. Keep the Second Schedule and guideline Table 6 side by side, with the PRIMS invoice as the final authority.
- Add the companion guidelines to the reading list. Alongside the main guideline, consult PRA/LPP/GUD/072, HPT/PER/GUD/041 and HPT/PER/GUD/042 at the corresponding project stages.
- Write the CSDT one tier above current requirements. Anticipate the tightening driven by the January 1, 2027 review milestone and the WHO ML3 effort.
- Keep version traceability. State the guideline number and version relied upon (HPT/PER/HTR/GUD/011 Rev.02) in submission documents, so differences can be traced when versions change.
- Check companion legislation in step. Section 1.1 of the guideline notes that the supply and use of devices must also comply with other applicable legislation (such as the Nuclear Regulatory Act); verify these in parallel at project kickoff.
5. Verification checklist
- □ Have the number and version of the guideline relied upon been recorded (HPT/PER/HTR/GUD/011 Rev.02)?
- □ Have the effective date of January 31, 2022 and the review date of January 1, 2027 been checked?
- □ Has the product been confirmed to fall within the scope of Section 1.3?
- □ Have the three companion guidelines been consulted (PRA/LPP/GUD/072, HPT/PER/GUD/041, HPT/PER/GUD/042)?
- □ Have fee calculations been checked against both the Second Schedule and guideline Table 6?
- □ Is a monitoring and comparison mechanism in place for guideline version changes?
- □ For drug-containing or combination products, has the track determination been completed first?
- □ Have parallel obligations under other applicable legislation (radiation, data protection, standards) been checked?
Regulatory information verified as of October 30, 2026.
In African markets, the GRZAN regulatory team focuses on jurisdictions including Kenya, Nigeria, South Africa and Egypt and maintains a verification mechanism built on primary regulatory sources: every regulatory conclusion published externally is based solely on official gazettes, the original text of agency guidelines and public legal databases, with the verification date stated; figures, timelines and amounts that cannot be located in official sources are marked as practical estimates, with no assumptions made.
GRZAN services cover: market access strategy and risk classification; preparation and review of registration dossiers (CSDT / STED / CTD formats); selection and management of Local Authorized Representatives (LAR); quality management system and GMP / GSDP compliance support; post-market vigilance, change management and recall response; import clearance and PVoC coordination; regulatory intelligence subscriptions and corporate compliance ledgers.
For a Kenya market access plan tailored to your product, contact the GRZAN regulatory team at info@grzanglobal.com.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us