Kenya Medical Device Regulatory Landscape: PPB, LN 100 of 2022 and Fees
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
Content verification date: October 30, 2026
1. Key takeaways
- Kenya’s device registration authority is the Pharmacy and Poisons Board (PPB), whose powers derive from the Pharmacy and Poisons Act (Cap. 244). The current rules, Legal Notice No. 100 of 2022, contain a zero-tolerance gate in Section 3: no person may import, manufacture or sell in Kenya any unregistered health product or technology.
- Kenya applies a four-tier risk classification (Class A, B, C, D) aligned with GHTF/IMDRF; Class A requires listing only, while Classes B, C and D require full evaluation. Fees are in US dollars: Class A USD 100, Class B USD 200, and Class C and Class D USD 1,000 each.
- The legal hierarchy gives statutes priority: the PPB guideline (HPT/PER/HTR/GUD/011, Rev.02) states in Section 1.1 that where the guideline conflicts with any written law, the written law prevails.
- PPB applies five evaluation criteria: safety, effectiveness, quality and performance, plus economic value (Section 5(1) of LN 100 of 2022) — a review dimension rare in the region, so pricing and value-for-money materials should be prepared in advance.
- Device market access is not a single PPB track: imports must also pass the Kenya Bureau of Standards (KEBS) Pre-Export Verification of Conformity (PVoC) and the Single Window import licence; devices containing radiation sources additionally require authorization from the Kenya Nuclear Regulatory Authority (KNRA).
2. Legal basis
| Instrument | Reference / number | Date of publication / entry into force | Provisions | Key points relevant to this article |
|---|---|---|---|---|
| Constitution of Kenya, 2010 | — | Promulgated 2010 | Article 43 | Right to health; cited in guideline Section 1.1 as the constitutional basis for device regulation |
| Pharmacy and Poisons Act | Cap. 244, Laws of Kenya | Enacted 1957 (current text per the Kenya Law database) | Sections 3A(a), 3A, 3B(2)(b), 3B(2)(i), 3B(2)(r), 44(1)(d) | Device guideline-making power; grant or withdrawal of marketing authorizations; cooperation with and reliance on foreign regulators; Cabinet Secretary rulemaking power |
| Pharmacy and Poisons (Registration of Health Products and Technologies) Rules, 2022 | Legal Notice No. 100 of 2022 | Made June 8, 2022; published at pages 684–709 of Kenya Subsidiary Legislation, 2022 (June 17, 2022 gazettement date is a practical estimate) | Sections 3, 4, 5(1), 5(4), 5(5), 10(1), 17 | Registration prohibition, applicant and holder qualifications, five evaluation criteria, five-year validity; repeals the 1981 Drugs Registration Rules (LN 147 of 1981) |
| Guidelines for Registration of Medical Devices including In-Vitro Diagnostics | HPT/PER/HTR/GUD/011, Rev.02 | Effective January 31, 2022; review date January 1, 2027 | Sections 1.1, 1.2, 1.3, 2.1 | Effect boundary of the guideline, authorization source, scope of application, risk classification principles |
| The Health Act, 2017 | No. 21 of 2017 | Assented June 21, 2017; effective July 7, 2017 | Sections 62–67 | Single regulatory authority for health products and technologies: functions, conditions, licences, standards, procurement |
| Nuclear Regulatory Act, 2019 | No. 29 of 2019 (now Cap. 243) | Assented December 23, 2019; effective January 10, 2020; Kenya Gazette Vol. CXXII—No. 4 | Section 5 | Establishes KNRA; repeals the former Radiation Protection Act (Cap. 243) |
| Standards Act | Cap. 496 | Current text per the Kenya Law database | — | Legal basis for KEBS functions and the PVoC programme |
| National Electronic Single Window System Act | Cap. 485D | Current text per the Kenya Law database | — | Legal basis for Single Window import licences |
3. Expert insights
Insight 1: Section 3 is a zero-tolerance gate, not a procedural reminder
Section 3 of LN 100 of 2022 is a prohibition: unless registered under the Rules, no person may import, manufacture or sell health products or technologies in Kenya. Three points matter. First, the prohibition covers import, manufacture and sale, so all three need registration support. Second, a health product under Section 2 expressly includes medical devices and diagnostics, so the view that devices need no registration does not hold in Kenya. Third, Section 17 repealed the 1981 Drugs Registration Rules, leaving no legacy stock.
Insight 2: A two-certificate structure — beyond registration, a Cap. 244 licence is required
Section 5(4) provides that PPB issues a registration certificate only if the applicant holds one of five valid licences under the Pharmacy and Poisons Act: the practice licence under Section 9A, the wholesaler licence under Section 27, the mining or agricultural poisons licence under Section 28, the Part II poisons sale licence under Section 32, and the manufacturing licence under Section 35A. For foreign manufacturers this pushes the duty forward to the Local Authorized Representative (LAR) — the LAR must itself hold a qualifying licence, or the product gets no certificate even after passing technical review.
Insight 3: Economic value is the fifth evaluation criterion, not a formality
Section 5(1) requires PPB to be satisfied on safety, effectiveness, quality, performance and economic value before registration. Section 5(2) further empowers PPB to approve or revise the sales name, labelling and promotional claims (including claims, route of administration, dosage, storage conditions, contraindications, side effects and precautions, pack sizes). GRZAN’s assessment: Kenya has written value for money into market access, so pricing position, comparisons with registered comparable products and health-economics materials should be prepared alongside clinical data.
Insight 4: Three external tracks run in parallel with PPB
Track one is standards and customs: under the Standards Act (Cap. 496), KEBS operates PVoC pre-export verification, and importers must first obtain a Certificate of Conformity (CoC). The second is the Single Window: after the CoC, batch PPB import licences are applied for via KenTrade, under Cap. 485D. The third is radiation: devices containing radiation sources require KNRA authorization under the Nuclear Regulatory Act, No. 29 of 2019, in addition to PPB registration. Guideline Section 1.1 also notes that registered devices remain subject to other applicable legislation.
4. Practical recommendations
- Classify first, then prepare the dossier. Make the question of whether a product is a medical device under Kenyan law the first project step, with a written determination record; for unclear boundaries (drug-device combinations, software, disinfection products, personal hygiene items), seek written clarification from the PPB medical devices unit early.
- Write the legal hierarchy into the internal compliance manual. Check the statutes first (Cap. 244 and LN 100 of 2022), then the PPB guideline; where they diverge, always follow the law and keep a written rationale.
- Prepare economic value materials in parallel. Prepare a short value argument page: target disease burden, existing alternatives, proposed pricing and supply stability, filed with the CSDT.
- Bring LAR due diligence forward. Before signing the authorization letter, verify that the local representative holds a qualifying Cap. 244 licence, is not already occupied by other product lines, and can handle vigilance and complaints.
- Build a timeline for all three tracks. Schedule PPB registration, PVoC, the Single Window import licence and KNRA authorization (if applicable) in parallel, so the certificate does not stall at customs.
- Track institutional change. The proposed KHPTA will take over PPB functions, and transitional provisions will decide how pending applications and existing certificates carry over; keep application numbers and acknowledgement receipts on record for each pending case (the Bill had not completed the legislative process as of the verification date — a practical estimate).
5. Verification checklist
- □ Has a written conclusion been obtained on whether the product is a device (including IVD) or a borderline product in Kenya?
- □ Has the risk class (A/B/C/D) been confirmed, with classification reasons mapped to the rules?
- □ Has it been confirmed that a foreign manufacturer must appoint a local authorized representative holding a qualifying Cap. 244 licence?
- □ Has the submission route (listing / full evaluation / abridged / fast-track / immediate) and dossier depth been confirmed?
- □ Have registration and retention fees been budgeted in US dollars (three tiers: USD 100 / 200 / 1,000)?
- □ Has a parallel timeline been set for PVoC, the Single Window import licence and KNRA authorization (if applicable)?
- □ Have economic value materials been prepared?
- □ Has a ledger of pending application numbers been kept, ready for verification when the regulator changes?
Regulatory information verified as of October 30, 2026.
GRZAN focuses on Kenya, Nigeria, South Africa, Egypt and other African jurisdictions. All external conclusions rely on official gazettes, agency guideline texts and public legal databases, with the verification date stated; unlocatable figures and amounts are marked as practical estimates.
GRZAN services cover: market access strategy and risk classification; preparation and review of registration dossiers (CSDT / STED / CTD formats); selection and management of Local Authorized Representatives (LAR); quality management system and GMP / GSDP compliance support; post-market vigilance, change management and recall response; import clearance and PVoC coordination; regulatory intelligence subscriptions and corporate compliance ledgers.
For a Kenya market access plan tailored to your product, contact the GRZAN regulatory team at info@grzanglobal.com.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us