Reprocessing of Single-Use Medical Devices: SAHPRA MD07-2025/26 Key Requirements

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

On 29 August 2025 SAHPRA published the communication Reprocessing of Single-Use Medical Devices (MD07-2025/26), regulating whether and how medical devices labelled as single use may be reprocessed. The core principle is that reprocessing is not a default-permitted activity: it may only be carried out within the regulatory framework and on the strength of validation evidence, and the reprocessor bears responsibility for the safety and performance of the reprocessed device once it is put back into use.

2. Legal basis

InstrumentReference / NumberPublication dateProvision
SAHPRA communication on reprocessing of single-use medical devicesMD07-2025/2629 August 2025Full text (conditions for reprocessing, validation and responsibility)
Regulations relating to Medical Devices and In Vitro Diagnostic Medical Devices (IVDs)Government Gazette No. 40480, Notice No. 1515Published on 9 December 2016Regulation 8(6) (compliance with the essential principles of safety and performance); Regulation 17 (vigilance); provisions on device housekeeping
Medicines and Related Substances ActAct No. 101 of 1965Enacted and published in 1965, subsequently amended; verified as at 29 October 2026Section 22C(1)(b); Section 14
SAHPRA guidance on essential principles of safety and performance of medical devices and IVDsSAHPGL-MD-01Current version, verified as at 29 October 2026Full text (essential principles that must still be met after reprocessing)

3. Expert insights

Insight 1: MD07-2025/26 completes a missing piece in South Africa’s device housekeeping framework. The 2016 Notice 1515 already contained device housekeeping content in its structure, and the 2021 draft went further by dedicating provisions to repair, servicing, refurbishment and the use of single-use devices. This shows that reprocessing has long been a concern of the South African system that had not been detailed, and the 2025 communication is its concretisation, giving operators a text they can finally design procedures around. The interval between the 2021 draft and the final communication also gave reprocessors several years of notice that these expectations were coming.

Insight 2: The reprocessor’s responsibility differs from that of the original manufacturer but is no lighter. Once a device labelled for single use is reprocessed and put into use, the reprocessor assumes responsibility for that device’s safety and performance, and must demonstrate conformity on three levels: essential principles conformity, risk management and vigilance reporting. This aligns with the logic, familiar from the EU MDR, that a reprocessor assumes manufacturer-like obligations for the devices it reprocesses. In practical terms this means the reprocessor needs its own technical file for the reprocessed device, not merely access to the original manufacturer’s documentation.

Insight 3: Validation evidence is the core of whether reprocessing can stand. Reprocessing must demonstrate that cleaning, disinfection or sterilisation, functional testing and packaging integrity have all been validated, and that a clearly justified upper limit on the number of reprocessing cycles exists. Practical estimate: the validation protocol and records should be established before reprocessing begins, not reconstructed afterwards when an inspection or an adverse event forces the question. Validation should also cover worst-case loads and the oldest allowable cycle, since those conditions define the safe envelope for reuse.

Insight 4: Reprocessing connects directly to the vigilance system. If an adverse event involves a reprocessed device, the reporting obligation falls on the reprocessor. The three-tier timelines of SAHPGL-MD-03 v4 (48 hours / 10 days / 30 days) apply equally. Reprocessing establishments must build their own vigilance procedures and cannot rely on the original manufacturer’s processes, which were never designed to cover their activities. Reprocessing establishments should therefore register as reporters in their own right and rehearse the escalation path before the first incident occurs.

Insight 5: Labelling and instructions for use must reflect the reprocessing status. Where a reprocessed device is no longer in its original factory single-use state, its labelling and instructions should reflect that fact truthfully. Practical estimate: where relabelling is involved, assess against MD10-2025/26 whether a licence amendment is triggered, since relabelling activities are expressly tied to the amendment route in South Africa. Where a device will be reprocessed repeatedly, the labelling should additionally state the maximum number of cycles allowed.

4. Practical recommendations

  1. Obtain the current text of MD07-2025/26 before starting reprocessing: build procedures to the communication’s requirements.
  2. Establish a complete reprocessing validation protocol and records: including the justification for cycle limits.
  3. Build your own vigilance procedures and reporting channels: the reprocessor is the responsible party.
  4. Ensure labels of reprocessed devices truthfully reflect their status: and assess amendment triggers.
  5. Integrate reprocessing into the quality management system: ISO 13485 coverage must include the process.
  6. Discuss technical feasibility with the original manufacturer: retain records of the communication.

5. Verification checklist

  • □ Current text of MD07-2025/26 (2025-08-29) obtained
  • □ Reprocessing validation protocol and records established (including cycle limits)
  • □ The reprocessor’s safety and performance responsibilities defined internally
  • □ Vigilance procedures and reporting channels built in-house (48 hours / 10 days / 30 days)
  • □ Labels of reprocessed devices truthfully reflect their status
  • □ Relabelling scenarios assessed against MD10-2025/26 for amendment triggers
  • □ Reprocessing processes covered within the ISO 13485 quality system
  • □ Records of technical communication with the original manufacturer retained

Regulatory information verified as at 29 October 2026. The regulations, guidance and effective dates described in this article are limited to the content published by the official sources listed under Legal basis as at that date; where official documents have been updated, the latest published version prevails.

  • South Africa local resource: GRZAN can assist manufacturers from the Chinese mainland and overseas in engaging the South African Health Products Regulatory Authority (SAHPRA) on medical device establishment licence (MDEL) applications and product registration preparation, including authorised representative (AR) scheme design, ISO 13485 certification route planning and CAB selection, GMDN coding and classification determination, technical documentation preparation, and the establishment of vigilance and adverse event reporting mechanisms.
  • Service capabilities: product attribute determination and risk classification (Class A/B/C/D), establishment licence application and renewal, reliance pathway justification, technical documentation and essential principles conformity argumentation, labelling and instructions-for-use localisation in English, and post-market vigilance and annual compliance information submission support.
  • Contact: hito.lin@grzan.cn; website www.grzan.cn.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail