Clinical Evaluation and PMCF in Saudi Arabia: MDS-REQ 1 Annex (6) Requirements
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Clinical evaluation in Saudi Arabia is governed by Annex (6) of MDS-REQ 1, titled Clinical Evaluation and Post-Market Clinical Follow-Up. Folding PMCF into the same annex shows that the SFDA treats clinical evidence as one continuous chain running from pre-market into the post-market phase, not as two separate reports. Existing foreign clinical data can be reused, provided the device to which the data correspond is identical to the device under application in design, manufacturing process, and intended use. This identity requirement shares one logic with item 2 of the 4 general eligibility criteria of the MDS-G030 reliance pathway, and once it is satisfied the review timeline can be compressed to 35 to 60 working days. Clinical trials conducted inside Saudi Arabia are additionally governed by MDS-REQ 2 (V4.1 / 2022-12-26). Manufacturers of high-risk devices should also consider preparing a Summary of Safety and Clinical Performance under Annex (9).
- Annex (6) merges clinical evaluation and PMCF into one evidence chain.
- The clinical evaluation report must reserve input slots for PMCF data, update frequencies, and re-evaluation triggers.
- Reuse of foreign clinical data turns on identity in intended use, design, and manufacturing process, not on the country of origin.
- Devices cleared by FDA 510(k) substantial equivalence are excluded from the reliance pathway, which interrupts the reuse channel.
- Clinical trials inside Saudi Arabia follow MDS-REQ 2 (V4.1 / 2022-12-26).
- Implantable devices require patient information and an implant card under Annex (8); high-risk devices should consider an Annex (9) summary.
2. Legal basis
The table below collects the instruments governing clinical evidence, their document numbers and dates, the relevant clauses, and the key points of the original text. Regulatory information verified as of October 29, 2026.
| Regulation | Document number | Publication date | Clause | Key point of the original text |
|---|---|---|---|---|
| Requirements for Medical Devices Marketing Authorization (MDS-REQ 1) | MDS-REQ-001-V6/211219 | V6.0 / 2021-12-19 | Annex (6) Clinical Evaluation and Post-Market Clinical Follow-Up | Evidence requirements for clinical evaluation and PMCF |
| Same instrument | Same number | Same date | Annex (9) Summary of Safety and Clinical Performance | Summary of safety and clinical performance |
| Same instrument | Same number | Same date | Annex (8) Implant Card and Information to be Supplied to the Patient with an Implanted Device | Patient information for implantable devices |
| Requirements for Clinical Trials of Medical Devices (MDS-REQ 2) | MDS-REQ-002-V4.1/221226 | V4.1 / 2022-12-26 | Full text | Requirements for clinical trials of devices conducted in Saudi Arabia |
| Medical Devices and Supplies Law | Royal Decree No. M/54 | 2021-02-18 (Hijri 1442/7/6) | Article 1 (Verification of Clinical Studies) | Definition of the verification of clinical studies |
| SFDA Reliance Approach for Medical Device (MDS-G030) | MDS-G-030-V1/260909 | V1.0 / 2026-09-09 | General Eligibility Criteria, item 2 | The device must be technically fully identical to the RRA-approved device in design, manufacturing process, and intended use |
| Depth thresholds of the clinical evaluation report under Annex (6) | No published SFDA breakdown text identified | — | — | Practical estimate: follow the current text of MDS-REQ 1 and GHAD review practice |
3. Expert insights
Merging clinical evaluation and PMCF in one annex changes the time horizon of the file. Placing clinical evaluation and post-market clinical follow-up in the same annex is not a layout choice. It defines clinical evidence as a continuous process: the pre-market evaluation produces an initial conclusion, and PMCF keeps feeding in evidence that corrects the conclusion in return.
For Chinese manufacturers the effect is organizational. The clinical evaluation report cannot be written as a document that is finished upon submission; it must reserve structural slots for PMCF inputs, naming the data sources, the update frequency, and the conditions that trigger re-evaluation. When GRZAN prepares a CER, the PMCF plan is normally given its own chapter with the trigger events for the next update stated explicitly.
Whether foreign clinical data can be reused depends on identity, not on the country of origin. MDS-REQ 1 does not restrict clinical data to studies performed in Saudi Arabia. The real gate is identity: the device to which the data correspond must remain consistent with the device under application in intended use, design, specifications, manufacturing information, and labeling. This wording runs fully parallel to item 2 of the MDS-G030 general eligibility criteria.
The workable route for reusing EU or US clinical data is therefore to complete the identity confirmation in the Rationale of the Reliance Form first, and then hang the same clinical evidence chain onto the Annex (6) structure. Conversely, if any FDA 510(k) device obtained clearance through substantial equivalence, it is explicitly excluded from the reliance pathway and the reuse channel for its clinical data is interrupted at the same time; see https://www.grzan.cn/regulations/saudi/why-fda-510k-excluded-from-reliance-14 .
Comparison with EU MDR and US FDA. The EU MDR requires the CER to argue in layers across the three sources of clinical data, the device itself, equivalent devices, and the literature, and imposes strict equivalence conditions; the Saudi Annex (6) is close to this structure and likewise brings PMCF inside. The US FDA does not use the CER as a universal carrier but requires clinical data separately for 510(k), De Novo, and PMA. China’s NMPA requires clinical evaluation or clinical trials for products outside its catalogue exempt from clinical evaluation.
The practical consequence for Chinese manufacturers is that a CER already written to EU MDR migrates to Saudi Arabia with the marginal cost concentrated on the Saudi-specific applicable requirements rather than on redoing the clinical argument. Companies holding only FDA 510(k) material must rebuild the argument, because the substantial equivalence route is explicitly excluded from the reliance pathway by MDS-G030 and the reuse channel for clinical data is interrupted with it. The relationship among the three major new rules appears at https://www.grzan.cn/regulations/saudi/saudi-2026-regulatory-update-review-49 ; there are 6 recognized regulatory authorities in total, and the eligibility details appear at https://www.grzan.cn/regulations/saudi/reliance-pathway-eligibility-and-rra-13 .
4. Practical recommendations
- Build the CER skeleton to Annex (6) with a separate PMCF plan chapter stating the update trigger conditions.
- When reusing foreign data, complete an identity comparison table covering intended use, design, specifications, manufacturing information, and labeling between the data device and the device under application.
- For clinical trials conducted in Saudi Arabia, prepare the ethics and approval materials in parallel against MDS-REQ 2 (V4.1 / 2022-12-26).
- For implantable devices, prepare the patient information documents and the implant card under Annex (8) at the same time.
- For high-risk devices, consider preparing the Annex (9) Summary of Safety and Clinical Performance to help reviewers locate key findings quickly.
- On the reliance pathway, cite the clinical evidence chain in the Rationale of the Reliance Form and leave room for SFDA requests for supplementary information.
- Serious adverse events or device deficiencies in clinical trials must satisfy the reporting requirements of MDS-REQ 2 and be captured in the NCMDR system.
- Have the GRZAN regulatory team check the current version of Annex (6) and confirm whether the depth thresholds for clinical evaluation have shifted (a practical estimate item).
- When filing several devices in one batch, complete the bundling classification under MDS-G028 first and only then judge whether the whole group qualifies for the reliance pathway; the order of the two decisions must not be reversed.
5. Verification checklist
- □ The CER has been built to Annex (6) with the PMCF plan as an independent chapter
- □ The identity comparison table for foreign clinical data is complete, covering intended use, design, specifications, manufacturing information, and labeling
- □ Whether the device falls within the reliance pathway exclusions has been confirmed (self-registration, substantial equivalence, custom devices, post-market safety issues, SaMD and AIeMD)
- □ Clinical trials inside Saudi Arabia have been tested for approval against MDS-REQ 2
- □ Patient information and implant cards for implantable devices have been prepared under Annex (8)
- □ A written judgment with reasons has been made on whether to prepare the Annex (9) summary
- □ The reporting route for serious adverse events in clinical trials has been written into internal procedures
- □ The current versions of Annex (6) and MDS-REQ 2 have been checked
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Saudi Arabian market. Our team prepares SFDA-compliant STAR dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us