Device, Medical Supply, Drug and Cosmetic Boundary in Saudi Arabia

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Characterising a boundary product in Saudi Arabia is a precondition, not a question resolved incidentally during submission. Article 1 of M/54 draws the boundary in two sentences: the principal intended action of a Medical Device is not achieved by pharmacological, immunological or metabolic means, although such means may assist it; and the Medical Supply covers medical materials or products used for diagnosis, treatment, replacement, orthopaedic support or disability situations, including medical gases. The updated Product Classification Guide adds that products claiming medicinal efficacy, or with ingredient levels exceeding prescribed limits, may be reclassified as drugs or health products, and cosmetics are limited to external use, without therapeutic substances or claims (a third-party account, practical estimate). Wrong characterisation writes off every subsequent investment, because the submission authority, technical file structure and review pathway all differ.

  • Article 1 of M/54 defines Medical Device (principal action not pharmacological, immunological or metabolic), Medical Supply (including medical gases), Assembled Medical Device and Innovative Medical Device.
  • Products claiming medicinal efficacy or exceeding ingredient limits may be reclassified as drugs or health products (practical estimate).
  • A surgical tray containing a drug requires the marketing authorisation for that drug issued by the SFDA drug sector, plus a component list letter.
  • An Assembled Medical Device may contain non-device items; the device characterisation of the set is unchanged.
  • MDS-G030 V1.0 application requirement 5 allows SFDA to request supplementary information for combination products, AIeMD and companion diagnostics.

2. Legal basis

The table collects the provisions that decide or affect characterisation at the device, medical supply, drug and cosmetic boundaries.

RegulationReference numberPublication dateArticles and key points
Law of Medical Devices and SuppliesRoyal Decree No. M/54; Hijri 1442/7/618 February 2021Article 1: definitions of Medical Device, Medical Supply (including medical gases), Assembled Medical Device and Innovative Medical Device
SFDA updated Product Classification GuideNo official reference number (third-party account)Published 20 January 2026, applicable from 20 November 2025 (practical estimate)Products claiming medicinal efficacy or exceeding ingredient limits may be reclassified as drugs or health products; cosmetics limited to external use
Guidance on Bundling Criteria for Medical Devices within a Single MDMA ApplicationMDS-G028 V1.0; MDS-G-028-V1/2605033 May 2026A surgical tray containing a drug requires the marketing authorisation for that drug issued by the SFDA drug sector; trays must submit a component list letter
SFDA Reliance Approach for Medical DeviceMDS-G030 V1.0; MDS-G-030-V1/2609099 September 2026Application requirement 5: SFDA may request supplementary information based on novelty, complexity and risk profile (AIeMD, companion diagnostics, combination products)
Guidance on Manufacturing Paths of Medical DevicesMDS-G011 V2.0; MDS-G-011-V2/2507077 July 2025Purposes such as custom devices may be exempt from MDMA (see MDS-REQ5)

3. Expert insights

Principal mode of action is the only pivot of the judgement

The M/54 definition of Medical Device uses a negative qualifier: the principal intended action is not achieved by pharmacological, immunological or metabolic means, although such means may assist it. This wording shares its origin with Article 2 of the EU MDR and the IMDRF definition, and leaves a channel for drug-device combinations: where the drug component acts only auxiliarily, the product remains a device, but the review treats it as a combination product. GRZAN’s order of analysis: list all mechanisms of action, rank them by contribution weight; if the drug’s metabolic action is the principal route to the intended use, take the drug channel; if the device’s physical barrier, support or delivery function is the main body, submit as a device with supplementary material on the drug component. The US FDA resolves the same dispute through the Office of Combination Products Request for Designation mechanism; China’s NMPA applies the principle that products acting mainly as devices are registered as devices, with a formal attribute determination procedure. No independent public attribute determination procedure has been seen in Saudi Arabia; as a practical estimate, GRZAN recommends a written e-GHAD classification decision as the characterisation evidence.

Medical Supply is a Saudi specialty and the category Chinese manufacturers most often omit

Article 1 defines the Medical Supply as medical materials or products used for diagnosis, treatment, replacement, orthopaedic support or disability situations, expressly including medical gases. In China such products may sit scattered under device, pharmaceutical packaging or industrial gas regimes, but in Saudi Arabia they are managed uniformly under the devices and supplies law and require an MDMA. Three omissions recur in GRZAN’s practice: medical gases (especially central gas supply systems), rehabilitation orthotic materials, and dressing products with a drug carrier function. Article 1 also defines the Assembled Medical Device — a set assembled to serve different user needs, which may contain non-device items. That matters: a non-device item in a kit does not change the device characterisation of the whole, but MDS-G028 V1.0 requires surgical trays to submit a component list letter itemising the component name, risk class, classification rationale, existing approval number, intended use, and the statutory manufacturer’s name and address.

The three highest-risk product groups: drug-device combinations, functional cosmetics, and health-claim products

The third-party account of the updated Product Classification Guide provides two clear red lines: products claiming medicinal efficacy, or with ingredient levels exceeding the prescribed limits, may be reclassified as drugs or health products; and cosmetics are limited to external use, without therapeutic substances or claims (practical estimate). Three groups carry the highest risk for Chinese exporters. First, drug-loaded dressings and drug-coated devices, such as antimicrobial silver-ion dressings and drug-eluting stents. Second, care products claiming repair, anti-inflammatory or healing-promotion effects: a therapeutic claim on the label or in promotional material may push the product into the drug or health product channel. Third, oral or transdermal health products, which cease to be devices once an ingredient level exceeds the limit. One workable self-check: compare all external claims (website, e-commerce, exhibition material) against the IFU and label in one pass — a single therapeutic claim anywhere triggers re-assessment of the characterisation. SFDA’s advertising requirements follow the IFU, and lay-user advertising must also be in Arabic.

The operating order for drawing the boundary is set out at https://www.grzan.cn/regulations/saudi/sfda-risk-classification-abcd-06 , and the kit component list requirements at https://www.grzan.cn/regulations/saudi/mds-g028-bundling-procedure-pack-and-ivd-29 .

4. Practical recommendations

  • Issue a written product characterisation judgement at project initiation: list all mechanisms of action, rank them by contribution weight, and state the device, drug or health product conclusion with reasons.
  • Base the characterisation on an e-GHAD classification decision; the application must submit detailed information on efficacy, mechanism of action and intended use (practical estimate).
  • Prepare drug-device combinations accordingly: the device part goes through the MDMA, the drug part is supported by composition, content, mechanism of action and safety evidence; where a surgical tray contains a drug, obtain the marketing authorisation issued by the SFDA drug sector.
  • Compare all external claims (website, e-commerce, exhibition material, social media) against the IFU and label in one pass, deleting or softening therapeutic claims; lay-user advertising must be in Arabic.
  • Check ingredient levels against the limits: where a level approaches or exceeds the prescribed limit, assess early the possibility of reclassification as a drug or health product (practical estimate).
  • Prepare the kit component list letter under MDS-G028, filling in the risk class, classification rationale, existing approval number, intended use, and the name and address of the statutory manufacturer for each component.
  • For sets containing non-device components, state the rationale under the Assembled Medical Device definition and explain in the technical file that the component does not affect overall safety and performance.
  • For combination products on the reliance pathway, reserve time for supplementary information — MDS-G030 expressly allows SFDA to request it based on combination product character.

5. Verification checklist

  • □ Product characterisation judgement issued and archived
  • □ Written classification decision obtained through e-GHAD (practical estimate)
  • □ All mechanisms of action listed and ranked by weight
  • □ Composition, content and safety evidence prepared for the drug part
  • □ Where the surgical tray contains a drug, the SFDA drug sector authorisation obtained
  • □ External claims compared against the IFU and label for consistency
  • □ Arabic versions prepared for lay-user advertising
  • □ Ingredient levels checked against limits (practical estimate)
  • □ Component list letter for kits prepared to the MDS-G028 fields

Regulatory information verified as of 29 October 2026.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Saudi Arabian market. Our team manages SFDA medical device marketing authorisations and Saudi Authorized Representative arrangements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN