Biocompatibility, Sterilisation and Shelf-Life Evidence

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Three areas of technical argumentation attract supplementary evidence requests more often than any others, and each maps to a defined group of points in Ek I. Knowing where reviewers concentrate their attention allows the file to be built right the first time, which in turn shortens the conformity assessment itself. The overlap is not accidental: properties, infection control and shelf life are the three fronts on which a device’s claim to be safe over its lifetime is argued, and reviewers sample them accordingly. The takeaways below identify where each argument must be built.

  • The three areas of technical argumentation most frequently drawing supplementary requests correspond respectively to three groups of points in Ek I: chemical, physical and biological properties (point 10), infection and microbial contamination (point 11), and the evidence relating to shelf life and the demonstration of performance over time.
  • Point 10 subdivides further: section 10.4 deals with CMR and endocrine-disrupting substances, and section 10.6 deals with nanomaterials. These two sections have become dense areas of review in recent years.
  • Class IIa and above devices typically enter the technical documentation assessment route (paragraph 4 point (a) of Madde 52 makes this explicit for Class IIb), so the completeness of the three groups of evidence above directly determines the length of the assessment.
  • The underlying logic is that the technical documentation must allow conformity to be assessed on its basis (paragraph 4 of Madde 10). Argumentation that states only conclusions, without methods and acceptance criteria, does not meet this standard.

2. Legal basis

The evidence requirements discussed here draw on two regulations and a short list of annex points: the general requirements on properties and on infection in Ek I, the assessment route for higher-risk classes in Madde 52, and the corresponding duties of IVD manufacturers. Because these points are read together during assessment, gaps in one area often draw questions into the others. The table consolidates the citations.

Name of regulationOriginal Turkish titleReference / numberPublication dateProvision
Medical Device RegulationTıbbi Cihaz YönetmeliğiResmî Gazete Sayı 31499 (1. Mükerrer)2021-06-02Madde 10 paragraph 4, Madde 52 paragraph 4 point (a); points 10, 10.4, 10.6 and 11 of Ek I
Regulation on In Vitro Diagnostic Medical Devicesİn Vitro Tanı Amaçlı Tıbbi Cihaz YönetmeliğiResmî Gazete Sayı 31499 (1. Mükerrer)2021-06-02Madde 50; Ek I

3. Expert insights

Insight one: point 10 is a set of requirements, not a single test item. It covers chemical, physical and biological properties, subdivided into sections such as 10.4 (CMR and endocrine disruption) and 10.6 (nanomaterials). Treating it as one biocompatibility report is not enough — what is needed is a set of documents covering the full materials argumentation. Each subsection raises its own questions about substances, exposure and justification, and a file that answers only the biocompatibility question leaves the other subsections open to challenge. A useful discipline is to read point 10 subsection by subsection against the product’s material list, marking each subsection as addressed, addressed with justification or not applicable, so the coverage decision is itself documented. Subsection coverage of this kind also gives the notified body a single place to see scope decisions, which shortens clarification rounds.

Insight two: sterilisation attaches to point 11, and shelf life attaches to the stability argumentation. Point 11 addresses infection and microbial contamination, and it is where the argumentation on sterilisation validation and sterile barrier systems converges. The shelf-life argumentation, for its part, must be built on stability data and packaging verification. The two topics travel together through review: a shelf life claimed without validated barriers, or a barrier validated without a claimed shelf life, reads as half of an argument that the reviewer expects to see completed. Planning the two argumentations together also avoids the scheduling conflict of running stability studies after the barrier design has been frozen. Reviewers tend to test the join between the two, so cross-references in the file should be explicit.

Insight three: the Class IIb technical documentation assessment is a sampling exercise, not a full inspection. The wording of paragraph 4 point (a) of Madde 52 is an assessment of at least one device in each device family group. This arrangement means that the unit selected must be complete enough, because its conclusions will be extrapolated to the whole group. Choosing the representative unit is therefore itself a strategic decision: it should be the configuration that exercises the fullest set of properties and justifications the family has to offer. Sampling therefore rewards preparation: the earlier the representative unit is agreed, the less risk that a late choice forces additional testing on an incomplete configuration.

4. Practical recommendations

  • For each group of evidence, build a four-part structure of conclusion plus method plus acceptance criteria plus index to raw data, benchmarked directly against the assessability standard of paragraph 4 of Madde 10.
  • For Class IIb and III, confirm in advance which unit or group the notified body will sample, and thicken the evidence for that group first.
  • Validation protocols relating to sterilisation and shelf life should be discussed with the notified body or a professional body before execution, to avoid rework afterwards.

5. Verification checklist

  • □ The evidence for point 10 of Ek I covers chemical, physical and biological properties.
  • □ Where CMR / endocrine-disrupting substances are involved, section 10.4 has been addressed.
  • □ Where nanomaterials are involved, section 10.6 has been addressed.
  • □ Sterilisation validation, sterile barriers and shelf-life stability data are complete (point 11 and the corresponding argumentation).
  • □ For Class IIb / III, the representative device to be sampled and its family grouping have been confirmed.

Regulatory information verified to 29 October 2026. The provisions and Ek I points listed in this article were checked against the current text and the annex on the Turkish legislative information system (MevzuatNo 38657).

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Turkish market. Our team prepares CE/UTS-compliant documentation, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail