Medical Device Labelling and Advertising Requirements in New Zealand
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Regulatory information verified as of 29 October 2026. New Zealand’s device labelling requirement sits in one provision, reg 12(4) of the Medicines Regulations 1984 (SR 1984/143): No person shall sell any medical device that does not bear the name of the manufacturer of the medical device or the name of the manufacturer’s distributor in New Zealand. It is the only substantive device-labelling requirement in the 1984 Regulations; the reg 12(5) exemption covers medicines only.
On penalties: the reg 64(1)(a) and (c) offence lists include neither reg 12 nor regs 4, 10, 17, 18 and 25, so breaching the device labelling, advertising and misleading representation duties produces no direct criminal penalty under that instrument. Do not invent an enforcement detour through reg 50: 50(1)(b) targets non-conformity with claimed specifications and (a) presupposes a Ministerial s 37 notice — neither fits a missing label. Only reg 62 and reg 50 (up to NZ$500) carry device penalties. Reg 62 covers claims that a device works by inducing, concentrating, channelling or producing magnetic, galvanic, electric, electronic, radiation or vibration forces or effects, or counteracting, shielding or protecting against them, with a written quantitative description to the purchaser and actual possession of those attributes — not a blanket provision for all electronic devices; breaches are offences under reg 64(1)(c). Labelling form: reg 17 requires prominent English, strong contrast, clarity and durability; reg 18(5) sets a 1.5 millimetre letter-height floor, reg 18(6) a 0.75 millimetre floor for small containers.
- Reg 12(4) is the sole device labelling requirement: manufacturer or New Zealand distributor name.
- Reg 12(5) exemption applies to medicines only.
- Only reg 62 and reg 50 are punishable on the device side, each capped at NZ$500.
- Letter height: 1.5 mm (reg 18(5)); 0.75 mm minimum for small containers (reg 18(6)).
2. Legal basis
The statutory basis for labelling and advertising of notified devices is set out below.
| Regulation | Official reference | Key dates | Relevant provisions |
| Medicines Regulations 1984 | SR 1984/143 | Made 5 June 1984; in force 1 August 1984; current version as at 10 July 2026 (Pae Ora terminology revision, not substantive) | reg 12(4) (device naming), reg 12(5) (exemption medicines only), reg 17, reg 18(5)(6), reg 25, reg 7, reg 10, reg 11(4), reg 50, reg 62, reg 64 (list (a) excludes regs 4(6), 10, 11(4), 12, 17, 18, 25; (b) withdrawal orders; (c) includes reg 62) |
| Medicines Act 1981 | 1981 No 118 (Public Act) | Royal assent 23 October 1981; in force from 1 August 1984; current version as at 10 July 2026 | Part 4 s 56 to 62 (s 56 advertisement includes labels and trade journals; s 57 7 prohibited matters; s 58 Schedule 1 diseases and panaceas; s 59(1)(2) true name and address required; s 59(4)(a)(b) and s 60 exemptions; s 59(5) breach up to NZ$1,000); s 61, s 63, s 66, s 98 |
| Medicines (Database of Medical Devices) Regulations 2003 | SR 2003/325 | Made 17 November 2003; in force 1 January 2004; current version as at 7 August 2020 | reg 10 (no implied official endorsement through database inclusion), reg 14 |
3. Expert insights
Insight 1: the only statutory labelling provision has no direct penalty. Reg 12(4) is a plainly worded sale prohibition, yet neither reg 64 offence list names it. The orders fall into two groups: withdrawal orders under (a) and (b) ((b) targeting devices not conforming with their claimed specifications; (a) presupposing a current Ministerial notice, with only s 37 available on the device side (s 35 concerns medicines only)), and disposal and destruction under (c) and (d), limited to what has been withdrawn under (a) or (b) and unable to self-trigger; (c) reaches only medicines, while devices fall under (d). Conclusion: reg 12(4) carries no direct criminal penalty and no alternative conviction route.
Consequences enter through the inspection power. Section 63(2) gives eleven limbs: (d)(e) inspection of articles and manufacturing or packaging processes have no threshold; seizure of labels or advertising ((j)) needs only a breach or non-conformity. Locked out are seizure of devices ((i)) and photography ((k)), which require reasonable belief an offence has been committed; breaching reg 12(4) is not an offence, so both are unavailable — a limited conclusion, not proof Medsafe is powerless. Criminal consequences come from s 66(3): refusal to provide records is an offence. More deterrent is s 98: the Director-General may publicly state a position on any device or related advertising, protected by qualified privilege — Medsafe’s leverage is often public naming. Note that s 79 extends liability but creates no offence: provisions outside the reg 64 lists yield no offence, so it cannot apply.
Insight 2: s 56 defines a label as an advertisement, so the two rulebooks connect. The s 56 definition of advertisement includes trade circulars, labels and trade journal advertisements; labels must not breach reg 25 or fall within the 7 prohibited matters in s 57. Medsafe states New Zealand has no statutory IFU requirement; once supplied, the IFU falls under reg 25: no comment contradicting, qualifying or varying label content, and no false or misleading statements, brands, pictures or marks. The s 56 publish definition carries catch-all limb (e) — bringing to the notice of the New Zealand public in any other manner — covering websites, e-commerce pages, social media and Chinese-language promotion aimed at New Zealand.
Insight 3: the Schedule 1 disease list is advertising’s hardest red line. Under s 58, medical advertising must not directly or by implication claim to prevent, mitigate or cure the Schedule 1 Part 1 diseases or conditions, including cancer, diabetes, arthritis, cataract, heart disease and infertility; nor claim panacea effects, absolute efficacy, or use or endorsement by practitioners, nurses or pharmacists.
Insight 4: s 59 is a duty, exemption and penalty trio. Section 59(1) requires medical advertisements to state the advertiser’s true name and address (a body corporate may substitute its registered office, s 59(2)); breach is punishable under s 59(5) by a fine not exceeding NZ$1,000. The duty is not unconditional: s 59(4)(a) disapplies s 59 where name and address disclosure regulations already apply, and s 59(4)(b) excludes exempted advertisers. Section 60 exempts advertisements distributed only to the s 58(1)(c)(ii) professionals or published in periodicals principally intended for them — professional journals qualify; internet pages and exhibition materials do not.
4. Practical recommendations
- Move the reg 12(4) naming requirement to packaging design: the first packaging specification must bear the manufacturer’s or New Zealand distributor’s name; relabelling after arrival costs far more.
- Validate size and durability at the proofing stage: the 1.5 millimetre floor in reg 18(5), the 0.75 millimetre floor in reg 18(6) for small containers, and the reg 17 form requirements.
- Screen copy against the s 58 Schedule 1 list: build the Part 1 disease names into a search-term list, compare website and social copy, and rewrite on any hit.
- Differentiate audiences to satisfy reg 10 and reg 11(4). Public-facing advertisements must, where applicable, state an accurate description, the purpose, precautions and contraindications; those directed at medical, dental, pharmacy and allied professionals must also state distribution restrictions, methods of use, effectiveness and limitations. Reg 11 was substantially amended on 1 August 2011 (SR 2011/245 reg 7).
- Ban official approval wording outright. Reg 7 of the 1984 Regulations and reg 10 of the 2003 Regulations jointly close off official endorsement phrases; marketing and distributor channels share one prohibited-word list.
5. Verification checklist
- □ Label bears the manufacturer’s or New Zealand distributor’s name (reg 12(4)); reg 12(5) exemption medicines only
- □ Letter height meets reg 18(5) at ≥1.5 mm, or reg 18(6) at ≥0.75 mm for small containers
- □ Accompanying graphics and instructions screened for reg 25 contradictions and false or misleading content
- □ External material contains no claim that breaching reg 12(4) is fined NZ$500, and no invented reg 50 detour
- □ Confirmed regs 4(6), 10, 11(4), 12, 17, 18 and 25 are outside the reg 64 lists; only reg 62 and reg 50 carry penalties on the device side
- □ Copy searched against the s 58 Schedule 1 list; advertiser’s true name and address stated (s 59, up to NZ$1,000)
- □ Public-facing advertisements include reg 10 items; professional-facing include reg 11(4) items
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the New Zealand market. Our team coordinates New Zealand sponsor arrangements, import requirements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us