Overseas Manufacturers and the New Zealand Sponsor Requirement

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

An overseas manufacturer cannot itself become the Sponsor (the New Zealand sponsor), nor complete WAND notification itself. The statutory definition appears in reg 3(a) of the Medicines (Database of Medical Devices) Regulations 2003 (SR 2003/325), and all three identities are prefixed by being a person in New Zealand: exporting or arranging exportation, importing or arranging importation, or manufacturing or arranging manufacture for supply. An overseas manufacturer cannot be the notifying party to WAND; the Sponsor must have a New Zealand entity presence. Where an overseas manufacturer supplies New Zealand directly, the New Zealand importer performing the import is the Sponsor. Medsafe’s official guidance states the same: Where an overseas company supplies directly to New Zealand, the importer is the sponsor.

The local-presence requirement is concrete and verifiable, not a statement of principle. Reg 5(1)(b)(i) of the 2003 Regulations requires the database information to include the address of the registered office or principal place of business in New Zealand of the sponsor, together with the manufacturer’s registered office or principal place of business address (whether in New Zealand or overseas) and contact telephone numbers or email addresses for both parties. A WAND record must therefore simultaneously show a New Zealand Sponsor address and a manufacturer address that may be anywhere.

The overseas manufacturer is not idle. Although not the Sponsor, it is the source of the information and evidence the Sponsor needs to perform: risk classification, GMDN codes, the unique product identifier (UPI) for Class III and AIMD devices, intended purpose, safety evidence, and technical support for recalls and adverse events.

2. Legal basis

The table below maps each matter onto the verified instrument, number and provision.

Sponsor definition and local presenceMedicines (Database of Medical Devices) Regulations 2003SR 2003/325; made 17 November 2003, in force 1 January 2004regs 3(a)(i)(ii)(iii) and the reg 3(b) exclusion
Database information (including the Sponsor’s New Zealand address)Medicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 5(1)(b)(i)(ii)(iii), reg 5(2) (UPI for Class III and AIMD)
Notification time limitMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 6 (within 30 working days of becoming Sponsor)
DeclarationMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 7
Information correctionMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 8 (10 working days)
Procedure dutyMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 9
PenaltiesMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 14 (breach of regs 6, 8, 9, 10; fine on conviction up to NZ$500)
Device safety evidential burden on importer or manufacturerMedicines Act 19811981 No 118; in force 1 August 1984ss 38(2)–(4) (45 day evidence window), s 38(8) (up to 6 months imprisonment or up to NZ$5,000)
Label must state manufacturer or New Zealand distributor nameMedicines Regulations 1984SR 1984/143; in force 1 August 1984reg 12(4)
Medsafe official guidanceMedsafe device definition pagehttps://www.medsafe.govt.nz/regulatory/DevicesNew/1Definition.asp, revised 15 January 2015Definition of a Sponsor

3. Expert insights

Insight 1: The reg 3(b) exclusion closes the appoint-an-agent route. Many overseas manufacturers first look for a New Zealand compliance services company to serve as the Sponsor. Reg 3(b) excludes persons acting on behalf of another person who is a New Zealand resident or carrying on business in New Zealand: if the services company acts purely as an import agent, the represented party is the Sponsor; if there is no such New Zealand represented party, the services company must itself actually perform or arrange the import before falling within reg 3(a)(ii). A nominal Sponsor therefore has no legal footing, and the Sponsor named in WAND records must be the party genuinely performing import, export or local manufacture arrangements.

Insight 2: Local presence lives in an address field, not in a certificate. New Zealand has no Sponsor registration certificate and no Sponsor qualification review. Presence is expressed through the reg 5 address and contact fields: the Sponsor must have a New Zealand registered office or principal place of business address and a reachable telephone or email. This differs from the EU authorised representative, which is registered with authorities, and the UK RP, which is held in MHRA records: in New Zealand, presence is recorded in notification data, not approved.

Insight 3: Multiple importers mean multiple Sponsors. Reg 3 defines the Sponsor by reference to the person, not the device. If the same product of the same manufacturer has two New Zealand importers, there are two Sponsors, each independently bearing the reg 6 notification duty and the reg 8 correction duty, each filing its own reg 7 declaration. The consequence for an overseas manufacturer’s channel management is direct: changing or adding distributors creates new notification duties and new record maintenance obligations.

Insight 4: The overseas manufacturer’s compliance role is information source plus technical endorsement. Medsafe issues no risk classification rulings; classification responsibility sits between Sponsor and manufacturer, with 22 rules in Schedule 2 and a judgement method that works backwards from rule 22 by exclusion. GMDN codes (IVDs use 5-digit codes) are supplied by the manufacturer. The s 38 safety evidence notice may be issued to the importer or the manufacturer, and evidence is due within 45 days of receipt. The overseas manufacturer’s practical role is therefore to supply accurate, verifiable classification, naming, UPI and intended purpose information to the Sponsor, and to carry the technical evidential burden when Medsafe opens a safety enquiry.

4. Practical recommendations

  • State Sponsor allocation and cooperation duties in distribution agreements. GRZAN recommends making explicit that the New Zealand importer is the Sponsor and completes WAND notification, and that the overseas manufacturer supplies risk classification bases, GMDN codes, UPI, intended purpose statements and change notifications. GRZAN also requires a named manufacturer contact, so the Sponsor is not left without an information source inside the 10 working day reg 8 window.
  • Create an immediate manufacturer-change-to-Sponsor-correction channel. Typical reg 8 triggers include manufacturer name changes, certificate lapses, and changes to risk classification or GMDN codes. Agree that the manufacturer gives written notice immediately on confirming a change, with the Sponsor then running its own 10 working day countdown (the specific notice time limit is a practical estimate).
  • Do not rely on nominal Sponsor arrangements. GRZAN’s reading of regs 3(a) and 3(b) is that a party not genuinely performing import, export or local manufacture arrangements can hardly hold Sponsor status under reg 3; once Medsafe or a purchaser traces responsibility, records and the actual responsible party will diverge.
  • Fix the channel list early. Every new importer generates an independent Sponsor and an independent notification. GRZAN recommends a channel policy under which supply to a new distributor is released only after its Sponsor status and completed notification are confirmed (this process design is a practical estimate).
  • Check labels in the same pass. Reg 12(4) of the 1984 Regulations requires the device to bear the manufacturer’s name or the name of its New Zealand distributor; the overseas manufacturer should confirm with the Sponsor how this field appears at the artwork stage, before leaving the factory.

5. Verification checklist

  • □ Written documents confirm that the New Zealand party performing import is the Sponsor and the overseas manufacturer does not act as Sponsor
  • □ The Sponsor’s New Zealand registered office or principal place of business address is confirmed as enterable in WAND (reg 5(1)(b)(i))
  • □ Contact telephone or email for both Sponsor and manufacturer is complete (reg 5(1)(b)(iii))
  • □ Nominal Sponsor and agent-held arrangements are excluded, with reg 3(b) assessed
  • □ Each New Zealand importer has separately confirmed Sponsor status and separately completed notification
  • □ The manufacturer has supplied risk classification bases and GMDN codes; UPI is supplied for Class III and AIMD
  • □ The manufacturer change notification mechanism is in the agreement, and the Sponsor’s internal 10 working day correction process is ready
  • □ Labels bear the manufacturer’s name or the name of its New Zealand distributor

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the New Zealand market. Our team manages Medsafe sponsor obligations and WAND entry requirements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail