Item Renewal in South Korea: The Five-Year Medical Device Certificate Cycle
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Article 49 of the Medical Devices Act, as amended by Act No. 17248 — promulgated April 7, 2020 and effective October 8, 2020 — established the comprehensive 5-year item renewal (품목갱신) system: all six certificate types, covering manufacturing licence, manufacturing certification, manufacturing notification, import licence, import certification and import notification, must be renewed every 5 years. The previous belief that one approval lasts indefinitely no longer holds from that date.
- Renewal is not a formalistic certificate exchange; the law sets three substantive gates. First, only devices actually manufactured or imported during the validity of the licence may apply for renewal. Second, items with serious safety or effectiveness problems are not renewed. Third, items for which documents are not submitted within the prescribed time are not renewed. Fees must also be paid, and devices exclusively for export and similar cases are subject to separate exceptions.
- On timing, note in particular: taking October 8, 2020 as the starting point of the 5-year cycle, the first batch of new certificates obtained after that date enters the first concentrated expiry window from October 2025. Shifting the 5-year cycle forward, the number of renewals to handle within 2026 will only increase. The common practice of submitting 180 days before expiry is marked in this guide as a practical estimate; the actual acceptance start point is subject to what the competent authority announces at the time.
2. Legal basis
The following instruments define the renewal cycle, the qualifying requirements and the supporting certificates.
| Medical Devices Act / 의료기기법 | Act No. 17248 (2020 amendment) | Amended and promulgated April 7, 2020, effective October 8, 2020 | Article 49 (item renewal: 5-year cycle, fee payment, actual manufacture/import requirement, non-renewal situations, exceptions such as export-only devices) |
| Medical Devices Act / 의료기기법 | Act No. 6909 (enacted) | Enacted and promulgated May 29, 2003, effective May 30, 2004 | Article 6 (manufacturing licence); Article 15 (import licence) — the source provisions of the six certificate types |
| Regulations on Approval, Certification and Notification of Medical Devices / 의료기기 허가·인증·신고 등에 관한 규정 | MFDS Notice No. 2022-52 | Effective July 29, 2022 | Item group classification, document composition of renewal applications and review methods |
| Standards for Medical Device Manufacture and Quality Management / 의료기기 제조 및 품질관리 기준 (KGMP) | MFDS Notice No. 2023-79 | 2023 | Mandatory for classes 2 to 4; certificate valid 3 years; validity must be confirmed at the time of renewal |
Note: the widely adopted practice of submitting 180 days before expiry is an industry operating position, not stated directly in the legal text, and is marked in this guide as a practical estimate; the fee amounts follow the current notices, and this guide does not list specific figures.
3. Expert insights
Insight 1: the renewal system clears dormant items out of the market — a mandatory health check for product line management.
Article 49 requires the renewal object to be a device actually manufactured or imported during the licence validity. The practical effect is that stockpiled items held with an approved certificate but no shipments do not qualify for renewal when they expire. GRZAN recommends using the first expiry window to run a product line inventory and converge the number of licences to the real business plan — every certificate retained adds one more set of 5-year-cycle fees and ancillary duties.
Insight 2: the three non-renewal situations are all concentrated eruptions of earlier management gaps.
Serious safety or effectiveness problems, late submission, and failure to meet the actual manufacture or import requirement look independent but all point to gaps in daily management. The third deserves emphasis: missing the prescribed submission time is itself grounds for non-renewal, meaning a missed window is not recovered by supplying one missing document — the application may have to start again from the first document.
Insight 3: the first concentrated expiry window stacks three timelines.
The 5-year item renewal cycle, the 3-year validity of the KGMP certificate, and the 10-day duty to modify device standard code (UDI) data after changes are usually owned by three different departments. When the three fall in the same year, missing any one affects the other two — an expired KGMP certificate directly drags on licence-type renewals, as described in the guide on the KGMP certificate; UDI data inconsistent with the licence information will trigger a demand for correction before the renewal review proceeds.
Insight 4: the renewal review looks not at the present but at what happened over the 5 years.
Foreign regulatory status, adverse events and change records all enter the field of view — exactly the material scope of re-evaluation under Article 9 of the Medical Devices Act, as described in the guide on the re-evaluation system. GRZAN’s judgement: preparing the renewal package as a small-scale re-evaluation is far safer than treating it as form filling.
4. Practical recommendations
- First take a full inventory of the six certificate types. List the certificates by the six categories of manufacturing/import crossed with licence/certification/notification, marking the effective date and expiry month of each, forming a single source of truth.
- Slice the renewal window into half-year segments. Use a point 6 months before the expiry month as the internal control start point, leaving buffer over the externally circulated 180-day position, and order submissions by document readiness.
- Use the first window to make product line trade-offs. For items with no actual import record during validity, assess the cost of simply giving up renewal rather than spending budget on products unlikely to launch.
- Align KGMP and UDI status at the same time. Mark each item’s KGMP certificate validity and UDI data update date on the same sheet, correcting any desynchronisation first, as described in the guide on the medical device standard code (UDI).
- Have the licence holder submit on a consolidated basis and retain receipts. Import-type renewals are executed by the Korean licence holder; state document provision deadlines and receipt-archiving duties in the agreement, and include completion evidence in the annual compliance file.
5. Verification checklist
- □ Legal basis and timing: verify Act No. 17248 (promulgated April 7, 2020, effective October 8, 2020); the deadline conclusion is traceable to Article 49 and the effective date is indeed October 8, 2020.
- □ Certificate inventory: check certificate by certificate across the six types; nothing is missing and every certificate has an effective date and expiry month.
- □ Actual manufacture/import records: sample shipment and import records within validity; every item applying for renewal has actual transaction flow during validity.
- □ Non-renewal risk: screen item by item against the three situations; no unresolved serious safety or effectiveness issues, and documents submitted within the prescribed time.
- □ KGMP status: check certificate validity and the periodic review application date; KGMP is valid for class 2 to 4 renewals and the application was filed 90 days before expiry.
- □ UDI data consistency: compare IMDIS data with licence information; changes were modified within 10 days and no inconsistencies remain.
- □ Submission lead time: check the internal start point against the practical estimate; scheduling runs backwards from the expiry month with lead time no shorter than the 180-day practical estimate.
- □ Evidence archiving: check submission receipts and fee payment vouchers; receipts and vouchers are complete and filed in the annual compliance file.
Regulatory information verified as of October 29, 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, Korean licence holder and local testing support for the Republic of Korea. Our team manages adverse event reporting and recall coordination with MFDS, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us