Substantial Equivalence Documentation: How to Select the Comparison Device
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Substantial equivalence (SE) in Korea is an element-by-element comparison exercise, with 6 comparison elements: intended use, mechanism of action (MoA), raw materials, performance, testing specifications, and instructions for use. The eligibility of the comparison device is a hard condition – it must be a product that has already obtained an item licence, item certification or item notification in Korea; an overseas registration certificate does not create comparison eligibility in Korea.
- Where all 6 elements hold, the product is treated as an equivalent device in the same category and may be exempt from parts of the technical documentation; where any element shows a significant difference, the product falls into the improved device or the new device category, and the documentation requirements rise accordingly. For in vitro diagnostic devices the comparison elements are 4: testing specifications and instructions for use do not apply.
- A failed comparison is not the end of the application. GRZAN handles failed comparisons through 3 remedial routes: replacing the comparison device, switching to the full technical documentation pathway, or supplementing performance data for the differing elements individually.
2. Legal basis
The instruments below set out the SE judgment criteria, the classification of new, improved and equivalent devices, and the IVD pathway that departs from the device rules. Regulatory information verified as of October 29, 2026.
| Regulation | Number | Date | Provisions |
| Regulations on Approval, Certification and Notification of Medical Devices | MFDS Public Notice No. 2022-52 | Effective July 29, 2022 | Substantial equivalence judgment criteria; classification of new, improved and equivalent devices |
| Medical Device Act | Act No. 6909 (enactment) | Enacted and promulgated May 29, 2003, effective May 30, 2004 | Article 2, Paragraph 2 (definition of the Technical Document), Article 6 (item licence), Article 15 (import item licence) |
| Enforcement Rule of the Medical Device Act | Prime Minister Decree No. 2127 | Partially amended and effective July 1, 2026 | Annex 1 (classification of medical devices and designated standards and procedures) |
| In Vitro Diagnostic Medical Devices Act | Act No. 16433 (enactment); Act No. 21525 (current) | Promulgated April 30, 2019, effective May 1, 2020; current version Act No. 21525, amended by other legislation April 7, 2026, effective October 8, 2026 | Class classification of IVDs and the licence / certification / notification pathways |
Scope of application of the 6 comparison elements. The table shows which elements apply to medical devices and to IVDs respectively, and where comparisons most often lose points.
| Comparison element | Medical devices | IVDs | Common weakness |
| Intended use | Applicable | Applicable | Use wording broader than the comparison device |
| Mechanism of action (MoA) | Applicable | Applicable | Confused with the principle of operation; mechanism argument lacks data |
| Raw materials | Applicable | Applicable | Material grade or supplier change not explained |
| Performance | Applicable | Applicable | Performance data from overseas methods, comparability not stated |
| Testing specifications | Applicable | Not applicable | Not mapped to the Korean item standard specifications |
| Instructions for use | Applicable | Not applicable | Inconsistent with the three Korean labeling layers |
3. Expert insights
Comparison device eligibility means obtained in Korea, not marketed overseas. Holding an NMPA registration certificate, an EU CE certificate or a US 510(k) clearance does not create comparison device eligibility in Korea. The comparison device must be a product that has already completed one of the three acts in Korea – item licence, item certification or item notification. The first task GRZAN performs at project initiation is to search Korean public information for usable comparison devices and build a candidate list, rather than writing the comparison table first and searching for a subject afterwards. The candidate list should name at least 2 to 3 products: the primary comparison device is used for the application, and the backup comparison device allows an immediate switch if the primary comparison fails.
Among the 6 elements, the first to collapse is usually not raw materials but instructions for use. Differences in raw materials are easy to identify and companies prepare explanations proactively. The instructions-for-use element tends to fail at the last moment: when the usage, frequency and applicable population wording in the Chinese leaflet is translated literally into Korean, it no longer matches the labeling scope of the comparison device, and a previously sound equivalence fails on the final element. The GRZAN approach is to finalize the instructions for use and the labeling texts before the comparison table.
IVDs carry only 4 elements; copying the device table of 6 creates two pieces of wasted work. The IVD comparison does not apply to testing specifications and instructions for use, because the analytical performance and clinical performance of IVDs are governed by separate requirements, see Article 29: Independent pathway for IVD devices and clinical performance studies. Applying the 6-element device table directly to an IVD forces comparability arguments on two inapplicable elements, which itself invites a deficiency request.
The three remedial routes after a failed comparison – the order of choice determines the cost. The first route is replacing the comparison device: find a closer model within the same item group, or narrow the scope of the application to align with the comparison device. The second is switching to the full technical documentation pathway: complete the dossier against the 7 elements described in Article 21: Technical Document structure and its correspondence with IMDRF STED. The third is supplementing performance data for the differing elements individually (practical estimate: whether the data suffice to bridge the difference must be confirmed with the accepting authority in advance). The cost of the three routes rises in that order, but the order of choice depends on where the difference lies: where the difference is in intended use or mechanism of action, replace the comparison device first; where it is in raw materials, complete the technical documentation first; where it is only in performance, attempt the single-element bridge first. GRZAN recommends exhausting the first route before moving on.
4. Practical recommendations
- Build the candidate list before writing the comparison table. Search Korean public information for similar products that already hold a licence, certification or notification in Korea, assemble 2 to 3 candidate comparison devices, and pre-score them against the 6 elements before fixing the primary comparison device.
- Give the mechanism of action its own section, supported by data. The MoA element must describe the action pathway clearly and cite verifiable literature or proprietary test data; a bare conclusion is not acceptable.
- Manage the comparison table as a controlled document. Bring the comparison table version number under the master technical documentation versioning, so it can be reused directly at change filings and item updates.
- Align the comparison scope with the item group scope. The models covered by the comparison table must match the item group division, and models across groups each need independent justification. If the comparison conclusion covers only part of the models in the group, the remaining models must still be compared individually in later applications.
- File the comparison conclusion together with the pathway determination. Whether the product is an equivalent device, an improved device or a new device directly determines the review pathway; complete the self-assessment before filing, see Article 20: Pre-market pathway selection decision tree.
5. Verification checklist
- □ Comparison device eligibility: search Korean item licence, certification and notification records; one of the three already obtained in Korea and verifiable.
- □ All 6 elements complete: verify element by element (4 elements for IVDs); all 6 (4 for IVDs) carry a conclusion, none left blank.
- □ Intended use consistency: compare the use wording word for word with the comparison device; not broader than the comparison device scope.
- □ Mechanism of action argument: check the mechanism description and supporting data; supported by literature or proprietary data, not a purely textual conclusion.
- □ Raw material comparability: check material grades and suppliers; differences explained and without impact on safety and effectiveness.
- □ Testing specification basis: query whether the item has a Korean standard specification; direct mapping where one exists, documented justification where none exists.
- □ Instructions for use consistency: compare against the finalized Korean instructions for use; consistent with the labeling scope of the comparison device.
- □ Classification conclusion: self-assess against the three classification criteria; conclusion clear and consistent with the intended pathway.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, Korean licence holder and local testing support for the Republic of Korea. Our team prepares MFDS-compliant technical dossiers, coordinates KGMP documentation, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us