NIDS and Designated Review Bodies: Class 1 and Class 2 Review in Korea

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Korea splits the review execution of Class 1 and Class 2 in two directions: the National Institute of Medical Device Safety Information (NIDS) accepts item notifications for Class 1 devices, with a statutory review period of 5 business days and an official fee of KRW 85,000; the designated review bodies undertake item certification for Class 2 devices, including technical document review and KGMP conformity audits, with 25 business days initially and 15 business days for a change.

The designated review bodies are designated by MFDS under Article 6-4 of the Medical Device Act; in practice the four bodies undertaking Class 2 review are KTL, KTR, KCL and KTC. Class 3 and Class 4 are outside this arrangement and are reviewed directly by MFDS, with 55 business days without clinical trials and 70 business days with clinical trials.

The value of understanding this boundary lies in scheduling: the Class 1 and Class 2 products of the same company run through two processes and two intake windows, and their correction rhythms do not affect each other. Managing them as one project distorts the timeline forecast.

2. Legal basis

The designation of review bodies and the division of work among institutions rest on the following instruments.

RegulationNumberPublication dateProvisions
Medical Device ActAct No. 6909 (enactment); No. 17248 (2020 amendment)Enacted and promulgated 29 May 2003Article 6-4 (designation of technical document review bodies), Article 6 (manufacturing licence), Article 15 (import business licence)
Enforcement Rule of the Medical Device ActPrime Ministerial Decree No. 2127Partially amended and promulgated 1 July 2026, effective the same dayAppendix Table 1 (standards and procedures for medical device classification and designation)
Regulations on Approval, Certification, Notification, etc. of Medical DevicesMFDS Notification No. 2022-52Effective 29 July 2022Provisions on technical document review and staged review
Standards for Medical Device Manufacturing and Quality Management (KGMP)MFDS Notification No. 2023-792023Mandatory for Classes 2 to 4, certificate valid 3 years, categories of conformity audits

Institutions, classes and division of work (supplementary table)

InstitutionUndertaken workStatutory review period
NIDSAcceptance of Class 1 item notifications5 business days
Designated review bodies (KTL, KTR, KCL, KTC)Technical document review and KGMP conformity audits for Class 2 item certificationInitial 25 business days; change 15 business days
MFDSReview of Class 3 and Class 4 item licencesWithout clinical trials 55 business days; with clinical trials 70 business days

3. Expert insights

Insight 1: NIDS and the designated review bodies are parallel assignments, not superior and subordinate.

The boundary of their work is set by the class; neither party re-examines the conclusions of the other. What GRZAN has encountered in projects is a company that submitted a correction question on a Class 2 product to NIDS and was told that this type of work is not accepted there, wasting the correction window. Correction replies must go back to the originally accepting body.

Insight 2: Class 2 products can shift to MFDS review in certain cases.

Class 2 by default follows certification by the designated review bodies, but where clinical data are needed, or the product involves digital medical products, review must shift to MFDS, and the statutory period and fee switch accordingly. GRZAN advises judging before filing whether any shift case exists, to avoid scheduling on 25 business days and then executing on 55 business days or more.

Insight 3: The KGMP conformity audit and item certification are two parallel tracks.

The KGMP certificate for Class 2 and above (valid 3 years) and the item certification proceed separately, but in the certification sequence the KGMP must come first. GRZAN advises drawing both tracks on the same Gantt chart and using completion of the KGMP as the milestone preceding acceptance of the item certification.

Insight 4: Choice of body is a project variable (practical estimate).

The law only provides that review is undertaken by bodies designated by MFDS; it does not give companies a free choice. The room for selection in practice comes from the timing of the application and the intake arrangement. GRZAN offers three criteria: first, whether the body’s concurrent schedule matches the product’s launch window; second, the body’s review experience with the item type; third, the responsiveness of correction communication. All three are practical estimates and must be judged against the actual situation at the time.

4. Practical recommendations

  • Fix the accepting body by class before back-planning the schedule. Schedule Class 1 on 5 business days, Class 2 on 25 business days and Classes 3 and 4 on 55 to 70 business days separately; do not manage a multi-class product line on one common cycle.
  • Route all correction communication only to the originally accepting body. Confirm the body name and contact person on the correction notice and avoid cross-institution enquiries.
  • Make the KGMP a precondition of item certification. Classes 2 to 4 must obtain a KGMP certificate valid 3 years, with periodic audits arranged before expiry.
  • Judge whether a shift-to-MFDS case exists. Situations requiring clinical data or involving digital medical products may shift review to MFDS; complete the judgement before filing.
  • Coordinate body scheduling for multiple filings in the same period. When several Class 2 items are filed in one year, confirm the intake arrangement with the licensed entity in advance and avoid cramming everything into the same window.

5. Verification checklist

  • □ Accepting body determination: match NIDS, the designated review bodies or MFDS by class, consistent with the statutory division table.
  • □ Screening for shift cases: check whether clinical data are needed or the product is a digital medical product; the screening conclusion is definite and documented.
  • □ Statutory review period: verify against the Appendix Table of the Enforcement Rule; the business-day basis is used.
  • □ KGMP sequence: check when the KGMP certificate was obtained; earlier than acceptance of the item certification.
  • □ Correction routing: check the accepting body named on the correction notice; identical to the originally accepting body.
  • □ Body scheduling coordination: check the intake arrangement for multiple items in the same period; not all crowded into one window.
  • □ Certificate validity: check the validity period of the KGMP certificate; 3 years, with a periodic audit scheduled before expiry.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, Korean licence holder and local testing support for the Republic of Korea. Our team manages MFDS premarket approvals and KGMP conformity, coordinates import requirements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail