Registration Changes and Prior Authorisation in Macao

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

After registration of a medical device in Macao, China, changes to the registration information fall into two categories: notification and prior authorisation. Article 20 of Administrative Regulation No. 11/2026 (changes to registration) provides that, for changes requiring prior authorisation, the decision period is up to 100 working days, while notifiable changes are handled under the notification procedure. Article 23(2) of Law No. 12/2025 is the overarching legal basis for this classification, with the specific requirements set out in Dispatch No. 11/ISAF/2026. The first task in change management is to determine which category a change belongs to — a misclassification leads directly to the compliance risk of implementing a change without approval, and may even affect the validity of the registration.

  • Changes to registration information are either notifiable or subject to prior authorisation.
  • The decision period for changes requiring prior authorisation is up to 100 working days.
  • Article 23(2) of Law No. 12/2025 is the overarching basis; Dispatch No. 11/ISAF/2026 details the requirements.
  • Implementing a change requiring prior authorisation without approval is a separate violation.
  • Change management must cover the entire registration dossier, not only the certificate particulars.

2. Legal basis

The change framework rests on the implementing rules, the parent law and the technical dispatch, with the fee schedule applying to certificate reissue. The table below lists the instruments and provisions cited in this article.

Regulatory instrumentReferencePublication dateProvisions
Implementing Rules of the Medical Devices Regulatory RegimeAdministrative Regulation No. 11/2026 (Official Gazette No. 20/2026)Published on 18 May 2026; effective on 1 July 2026Article 20 (changes to registration; decision period for changes requiring prior authorisation up to 100 working days); Article 19 (reissue of the registration certificate); Article 13 (review of registration, formal review 20 working days)
Medical Devices Regulatory RegimeLaw No. 12/2025 (Official Gazette No. 30/2025)Published on 28 July 2025; effective on 1 July 2026Article 23(2) (institutional basis for changes to registration information); Article 13 (obligations of registration holders)
Technical Requirements and Specific Rules for Changes to Medical Device Registration InformationDispatch No. 11/ISAF/2026Published in 2026; effective on 1 July 2026Full text of the dispatch
Dispatch of the Chief Executive No. 144/2026 approving the Fee Schedule for Medical Device Business Activities and Medical Device RegistrationDispatch of the Chief Executive No. 144/2026 (Official Gazette No. 24/2026)Published on 15 June 2026; effective on 1 July 2026Table 2 (registration renewal MOP 200; reissue of the registration certificate MOP 200)

3. Expert insights

Insight one: 100 working days is a maximum decision period, not an average duration. The wording is up to 100 working days, comparable in magnitude to the registration decision periods of 70 (Class IIb) and 100 (Class III) working days. The intensity of the review of a change is not necessarily lower than that of the initial registration — especially for substantive changes affecting technical requirements, intended use or manufacturing processes. Practical estimate: schedule substantive changes on the basis of 100 working days plus 20 working days of formal review; do not assume that a change is faster than a registration.

Insight two: the dividing line between notification and prior authorisation lies in the impact on safety and effectiveness. Dispatch No. 11/ISAF/2026 enumerates which changes must be notified and which require prior authorisation. The underlying criterion is that matters not affecting the safety and effectiveness of the device go through notification, while matters that may affect them go through prior authorisation. Where classification is unclear, companies should consult the ISAF rather than default to the lighter category — this is the lowest-cost form of risk control.

Insight three: changes and renewal interact with each other. Registration is valid for 5 years, with the renewal window falling 90 to 180 days before expiry. If a major change is submitted close to expiry, the change review and the renewal review may run in parallel or conflict. It is advisable to schedule major changes in the first and middle parts of the certificate validity period, away from the renewal window.

Insight four: implementing a change without authorisation is a violation in its own right. Law No. 12/2025 establishes a system of administrative and criminal liability for violations of the regime. Changing first and reporting later is tolerated in some scenarios in the Chinese mainland, but under the legal framework of Macao, implementing a change requiring prior authorisation without obtaining it is itself a violation. Mainland enterprises localising into Macao need focused training on this point.

Insight five: change management must extend beyond the particulars printed on the certificate. Registration information is not limited to the few items on the certificate; risk analysis documents, technical requirements and test summary reports, clinical evaluation documents and quality management system documentation are all part of the registration dossier. Any substantive update to these documents may trigger the change procedure. Companies should maintain a dossier-to-change linkage list, rather than watching the certificate only.

4. Practical recommendations

  • Build a change classification matrix: determine notification versus prior authorisation first, then run the procedure.
  • Reserve 120 working days for major changes (20 working days of formal review plus 100 working days of decision period).
  • Keep major changes away from the renewal window: schedule them in the first and middle parts of the certificate validity period.
  • Prohibit change-first, report-later practice: write the prohibition into local SOPs and train teams from the Chinese mainland.
  • Archive change records together with the dossier: create a paper trail for the next renewal and inspection.

5. Verification checklist

  • □ Determined whether the change is notifiable or requires prior authorisation
  • □ Changes requiring prior authorisation scheduled on the basis of up to 100 working days
  • □ Formal review of 20 working days built into the schedule
  • □ Change list of Dispatch No. 11/ISAF/2026 checked item by item
  • □ Confirmed that the change avoids the renewal window (90 to 180 days before expiry)
  • □ Red line of not implementing changes without authorisation embedded in internal SOPs
  • □ Change documentation updated and archived together with the registration dossier
  • □ Changes affecting labelling and instructions localised in parallel (Chinese, Portuguese, English)

Regulatory information verified as of 29 October 2026. The laws, guidelines and effective dates described in this article are limited to the content published by the official sources listed under Legal basis as of that date; where official documents are updated, the most recently published version prevails.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for Macao, China. Our team coordinates local licence holder arrangements, import and supply requirements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail