IVD-Specific Documentation Requirements
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Acuerdo No. 0631-ARSA-2023, the Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario, spreads the documentation requirements for in vitro diagnostic devices (IVDs) across three places: Article 32, item 6 (at the risk classification stage) requires six items — the technical datasheet (ficha técnica), the insert (inserto), storage conditions, storage and transport conditions, the safety data sheet (hoja de seguridad), and the qualitative and quantitative formulation expressed in measurement units, percentages or ingredients, with CAS numbers; Article 43, Section 2 (at the registration stage) requires six items — technical datasheet, insert, technical documentation supporting shelf life or stability, storage and transport conditions, safety data sheet, and qualitative and quantitative formulation (by measurement unit, dosage, percentage or ingredient); and Article 44, item 3 (from Class IIA) adds the validation report of the test, assay or product. The GRZAN regulatory team notes: the IVD lists at the classification and registration stages are not identical — the registration stage adds the stability document; as a practical estimate, prepare the full set once to the registration standard and reuse it at classification, saving one round of supplements.
2. Legal basis
The table below maps each IVD requirement to its provision and key point.
| Level | Instrument | Provision | Key point |
|---|---|---|---|
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 32, item 6 | IVD six items at the classification stage: technical datasheet, insert, storage conditions, storage and transport conditions, safety data sheet, qualitative and quantitative formulation (with CAS numbers) |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, Section 2 | IVD six items at the registration stage: technical datasheet, insert, technical document supporting shelf life or stability, storage and transport conditions, safety data sheet, qualitative and quantitative formulation |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 44, item 3 | From Class IIA, a validation report of the test, assay or product is added |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 3, Section II | Definition of Dispositivo Médico de Diagnóstico In-Vitro (IVD); definition of Accesorio para un Dispositivo Médico de Diagnóstico In-Vitro |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 84, Section 1 o) | The primary label must carry the IVD use statement (declaración de uso de diagnóstico in vitro o IVD) |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 18 | Transporting IVDs that require special transport conditions requires a transport licence |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 46, item 5 e) | The Class III equivalent documents include the IVD CE certificate issued by a European Union notified body (98/79/EC) |
3. Expert insights
Insight 1: the two lists differ by exactly one item — the stability document. Article 32, item 6 (classification), lists condiciones de conservación and almacenamiento y transporte side by side; Article 43, Section 2 (registration), replaces the former with documento técnico que respalde la vida útil o estabilidad. In other words, the registration stage demands evidence of stability, not merely a statement of storage conditions. Preparing the full set once to the registration standard is the simplest approach.
Insight 2: the safety data sheet maps to chemical management, not clinical evidence. Article 32, item 6 e), and Article 43, Section 2 e), both require the hoja de seguridad (safety data sheet, SDS). For IVDs containing chemical reagents, this document is the basis for transport, storage and emergency handling, and it sits on the same logical chain as Article 18, which requires a separate transport licence for IVDs needing special transport conditions.
Insight 3: the validation report is added from Class IIA. Article 44, item 3, requires IVDs to submit informes para la validación de la prueba, ensayo o del producto — reports for the validation of the test, assay or product. This is an important dividing line: Class I IVDs need no validation report; Class IIA and above do. When planning IVD products, classify under Article 29 first, then decide whether validation studies are needed.
Insight 4: the IVD qualitative and quantitative formulation must carry CAS numbers. Article 32, item 6 f), requires the formulation con su número CAS — with its CAS number — echoing the CAS number field in Article 32, item 1 o). For multi-component reagents this means a CAS number for every chemical component, not only the main ingredient.
Insight 5: the primary label must carry a dedicated IVD use statement. Article 84, Section 1 o), requires the IVD primary label to include the declaración de uso de diagnóstico in vitro o IVD. This is a mandatory label field triggered by the product category, operating in the same way as the sterile symbol triggering ISO documentation (Article 43, item 13): whatever appears on the label must have a corresponding supporting document or statement.
4. Practical recommendations
- Prepare the IVD set once to the six-item registration standard and reuse it at the classification stage.
- Prepare stability evidence as a formal technical document, not merely a statement of storage conditions.
- For IVDs containing chemical reagents, complete the safety data sheet in step with the transport licence application.
- Schedule validation studies early for Class IIA and above; do not schedule them for Class I.
- Provide a CAS number for every component of multi-component reagents, not only the main ingredient.
- Add the IVD use statement field to the primary label and update the insert at the same time.
5. Verification checklist
- □ The IVD documentation set is complete once, to the six-item registration standard
- □ Stability evidence has been prepared as a formal technical document
- □ IVDs containing chemical reagents have a safety data sheet in place
- □ Validation studies are scheduled by class (from IIA)
- □ CAS numbers are provided for every component of multi-component reagents
- □ The primary label includes the IVD use statement field
Related reading: [Article 21 — The sixteen Class I dossier items](/honduras-medical-device-registration/class-i-dossier-sixteen-requirements/) | [Article 22 — What Class IIA adds](/honduras-medical-device-registration/class-iia-iso-13485-and-software/) | [Article 34 — The transport licence](/honduras-medical-device-registration/transport-licence-article-18/)
Regulatory information verified as of October 29, 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Honduran market. Our team manages Secretariat of Health (SESAL) registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us