Separate Registration and New Registration Triggers

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Costa Rica has two start-over triggers that are easily mixed up, set out in Articles 10.8 and 10.9 on the one hand and Article 10.5 on the other: first, EMB that are similar but differ in composition must be registered separately; second, where a change of the specific use, that is the indication for use, alters the risk classification, a new registration is required. At the same time, Article 10.5 treats a product change that alters the risk classification as a new product, requiring a fresh application under Articles 11, 12 or 13. Distinguishing these two triggers, and distinguishing them from the post-registration change channel of Article 18, is the core of change management in Costa Rica.

2. Legal basis

The table below sets out the separate registration and new registration triggers alongside the change provisions that apply when the classification does not change.

LevelInstrumentProvisionKey point
Technical regulationRTCR 505:2022Article 10.8EMB that are similar but differ in composition must be registered separately
Technical regulationRTCR 505:2022Article 10.9A change of specific use (indicación de uso) that alters the risk classification requires a new registration
Technical regulationRTCR 505:2022Article 10.5A product change that alters the risk classification is treated as a new product and must be re-registered under Articles 11, 12 or 13
Technical regulationRTCR 505:2022Article 10.6Modifications, reductions or additions to packaging sizes are handled as post-registration changes under Article 18.10
Technical regulationRTCR 505:2022Article 18.14Use changes or extensions: a medical specification containing the new use must be submitted; Clase 3 and 4 must also submit clinical evidence
Technical regulationRTCR 505:2022Article 18Post-registration changes must be approved before implementation, except for notification-type changes
Technical regulationRTCR 505:2022Article 8.1Where a product could fall into several classes, it is classified in the higher risk class

3. Expert insights

Insight 1: Article 10.8 and Article 10.7 are neighbouring provisions and are complete only when read together. Article 10.7 permits seven forms of grouping, and Article 10.8 immediately afterwards provides that products differing in composition must be registered separately. This adjacent structure shows that the legislator deliberately drew a boundary while offering the grouping convenience: the legitimacy of grouping comes from identity of use and structure, and a difference in composition cuts grouping off directly. In practice, the judgement on whether products differ in composition should be made on the formulation and material composition, not on the product name or the model designation.

Insight 2: The division of labour between Article 10.5 and Article 10.9 is that the former governs product changes and the latter governs use changes, and the common threshold for both is a change of risk classification. The consequence of both is the same, re-registration; the difference lies in what triggers them. Conversely, where a change does not alter the classification, no new registration is triggered and the post-registration change route of Article 18 applies: packaging size movements go through Articles 10.6 and 18.10, and use extensions go through Article 18.14, although use extensions for Clase 3 and 4 still require clinical evidence to be submitted separately.

Insight 3: Article 18.14 reveals an intermediate zone that is easy to misjudge: a use extension is itself a change, but it can slide into re-registration. Article 18.14 permits use changes or extensions to be handled as post-registration changes, conditioned on submitting a medical specification that contains the new use, and for Clase 3 and 4 on submitting clinical evidence supporting the new use, pursuant to Article 12.2. But Article 10.9 simultaneously provides that a use change that alters the classification requires a new registration. The correct sequence is therefore to first assess, under Article 8.1 and Annex A, whether the new use changes the classification; if it does, take the re-registration route, and only if it does not take the Article 18.14 route.

Insight 4: The first sentence of Article 18 requires changes to be approved before implementation, which forces the assessment to happen up front. The source text reads that the applicant must tramitar su aprobación a la autoridad reguladora nacional previo a su implementación. Combined with Article 10.5, if a company implements a change without assessment and that change alters the classification, two consequences are triggered at once: implementing a change without approval under Article 18, and selling a new product without registration under Article 9, the latter falling into the serious infringement of Article 2, item II.b) of the approving decree.

4. Practical recommendations

  • Establish a change assessment checklist: for any movement in formulation, materials, structure or use, first determine whether it alters the risk classification under Article 8.1.
  • Register products with different compositions separately under Article 10.8, and do not merge them under the name of a family or system.
  • Plan changes that alter the classification directly as new registrations under Articles 11, 12 or 13, and do not route them through the Article 18 change channel.
  • Handle use extensions that do not alter the classification under Article 18.14, and prepare the corresponding clinical evidence by class, mandatory for Clase 3 and 4.
  • Complete the filing of all changes before implementation as required by the first sentence of Article 18, and treat implementing changes before approval as an internal red line.

5. Verification checklist

  • □ Assessed whether each change alters the risk classification (Articles 8.1, 10.5 and 10.9)
  • □ Products with different compositions registered separately (Article 10.8)
  • □ Changes that alter the classification planned as new registrations (Articles 11, 12 or 13)
  • □ Use extensions that do not alter the classification prepared under Article 18.14
  • □ Clinical evidence prepared for Clase 3 and 4 use extensions (Article 12.2)
  • □ All changes filed before implementation (first sentence of Article 18)

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Costa Rican market. Our team manages Ministry of Health (Ministerio de Salud) sanitary registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail