New Supplier Information and Documented Procedures Requirements for MDEL Applicants

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

From December 14, 2026, MDEL applicants must provide two new items of information: new paragraph 45(f.1) requires the name and address of any person, other than a manufacturer, from whom the applicant has purchased devices for import or for distribution purposes; new paragraph 45(f.2) requires, for each such person, the class or classes of devices that the applicant is importing or distributing from them. At the same time, five documented procedure obligations change from a declaration into express requirements of the Regulations.

  • Supplier information is captured at the level of name, address and device class, per supplier.
  • The new items apply to every supplier except manufacturers, and must cover all suppliers in the applicant’s chain.
  • Supplier information becomes an annually refreshed item through the amended ALR under paragraph 46.1(1).
  • Section 58.1 is an entirely new provision covering six activities for Class II to IV devices, including corrective actions.
  • Sections 59(3) and 59(4) layer the reporting procedure duties differently for Class I and for Class II to IV devices.

2. Legal basis

The new requirements come from discrete items of SOR/2026-110, each with its own section of the amended Regulations. The table below lists the items, what each adds or replaces, and the common effective date.

Regulation or documentProvisionsDatesSource
SOR/2026-110, item 2(2)Adds paragraph 45(f.1) (supplier name and address) and 45(f.2) (device classes per supplier)Effective 2026-12-14Canada Gazette, Part II
SOR/2026-110, item 4Adds section 52(3): distribution record proceduresSameSame
SOR/2026-110, item 5Replaces section 58: establish, implement and maintain documented proceduresSameSame
SOR/2026-110, item 6Adds section 58.1: procedures for handling, storage, delivery, installation, servicing and corrective actions for Class II to IVSameSame
SOR/2026-110, item 7Adds sections 59(3)(4): preliminary and final report procedures for Class I and Class II to IVSameSame
SOR/2026-110, item 8Adds section 61.2(4): procedure duties for the provision of informationSameSame

Because every item shares the same effective date, the compliance work for the supplier information and the five procedure duties can be planned as one programme with a single deadline, even though the underlying provisions address different activities across the product lifecycle.

3. Expert insights

Insight 1: the granularity of supplier information is the class, not the individual device.

New paragraph 45(f.2) requires, for “each person referred to in paragraph 45(f.1), other than a manufacturer, the class or classes of devices that the person is importing or distributing from that person”. This lowers the granularity required for maintaining the information: it is maintained by class rather than device by device, but it must still cover every supplier. For a company with a long supplier list, that difference is material to the maintenance workload, since a single supplier may serve an entire class of products without the company needing to enumerate each catalogue item underneath.

Insight 2: supplier information must be refreshed at the ALR; it is not a one-time submission.

In combination with the amended paragraph 46.1(1), under which the ALR must include updates to the section 45 information and documents, supplier information becomes an item that is refreshed every year. Companies should set up an annual review mechanism for their supplier list, with named ownership, so that changes in the supply chain during the year are collected continuously and reconciled once at the review rather than reconstructed from memory when the deadline arrives.

Insight 3: section 58.1 is an entirely new provision covering six activities.

The new section 58.1 applies to Class II, III or IV devices and requires manufacturers, importers and distributors each to establish, implement and maintain documented procedures for, as applicable to them, the handling, storage, delivery, installation and servicing of devices, and for corrective actions taken in respect of devices. This is the complete list of six activities. The word each matters: the duty attaches separately to every licence holder in the chain according to the activities it actually performs, so each company must scope the six activities against its own role rather than relying on another party’s procedures.

Insight 4: sections 59(3) and 59(4) are layered by device class.

Section 59(3) targets manufacturers and importers of Class I devices (covering preliminary and final reports under subsections 59(1) and (1.1), including foreign events); section 59(4) targets manufacturers and importers of Class II to IV devices (covering reports under subsection 59(1)). The scope of the required procedures differs between the two groups, so a company handling devices of different classes must establish the respective procedures separately rather than assuming one reporting procedure covers both regimes.

4. Practical recommendations

  • Build a supplier list with three columns: supplier name, address and device class, reviewed annually.
  • Compare existing documentation against each of the five procedures: distribution records (52(3)), complaint handling and others (58), the six Class II to IV activities (58.1), event reporting (59(3)(4)), and provision of information (61.2(4)).
  • Confirm applicability activity by activity for section 58.1, and do not omit corrective actions.
  • Establish the Class I and Class II to IV event reporting procedures separately (sections 59(3) and 59(4)).
  • Complete remediation before December 14, 2026 with operating records in place, not just documents on a shelf.

5. Verification checklist

  • □ The supplier list with three columns (name, address, device class) is established
  • □ Supplier information is included in the annual ALR refresh mechanism
  • □ The five documented procedures have been compared and completed one by one
  • □ Applicability of the six activities in section 58.1 has been confirmed activity by activity
  • □ The Class I and Class II to IV event reporting procedures are established separately
  • □ Remediation is scheduled to complete before December 14, 2026

Regulatory information verified as of October 28, 2026.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Canadian market. Our team manages Health Canada Medical Device Licence and MDEL applications, prepares risk-classified dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail