Bangladesh SaMD: Software as a Medical Device

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

In Bangladesh, software is a medical device, and there is no separate SaMD channel. The Drug and Cosmetics Act, 2023 (Act No. 29 of 2023, in force 2023-09-18) lists software directly among the device forms, so software as a medical device follows the same four-class scheme and the same two-step registration process as hardware. Class B, C and D products require notified body QMS certification in the same manner, and Class C and D products require a reference country CFS in the same manner. Nothing in the framework creates a lighter track for products whose only physical manifestation is a download or a hosted service. For software developers the question has therefore shifted from whether the product is regulated to which class it belongs to and what evidence that class demands.

  • Software appears in the statutory list of device forms, without qualification.
  • No exemption exists for standalone software or for low-risk software.
  • Classification turns on the information purpose the software serves, not on its delivery model.
  • Digital products still enter a paper-based submission system.
  • Data location questions sit outside the guidelines and remain a pending verification item.

2. Legal basis

The statutory hook is a single word in the definition clause, and everything else follows from it. The table sets out the provisions relied on and flags the one row that rests on a single source rather than on the guidelines.

MatterLegal sourceProvision
Date in force and act numberDrug and Cosmetics Act, 2023Act No. 29 of 2023, in force 2023-09-18
Software listed in the device definitionSame ActDefinition clause (a): software
Six intended purposesSame ActDefinition clause (a)(i)-(vi)
Computer-simulated diagnostic reagentsSame ActDefinition clause (a): in silico reagent
Four-class scheme and conformity assessmentDGDA Medical Device Registration Guidelines (Device Classification v 1.0)§4.4, §4.5, §4.6
Standards appliedSame guidelines§7.1
A data localisation explanation may be requested where servers sit abroad; cloud-based AI-assisted diagnosis requires additional data processing and storage documentationRegistration practice compilation (pending verification, single source)Not applicable

3. Expert insights

Insight one: software entered the statute recently, but the drafting is concrete rather than aspirational. The law that took effect in September 2023 names software in the definition, and in the same clause it names in silico reagents. The two together cover, at the level of legislation and in a single instrument, both standalone software and model-based diagnostic output. That closes the conventional defence that an algorithm service is not a device. A manufacturer cannot describe its product as a service layer, a decision-support tool or an analytics platform and expect the classification question to disappear; the statutory text already answers it. The argument has to be made on purpose and output, not on delivery technology.

Insight two: the class of a SaMD product resolves to its information purpose. Definition clause (a)(vi) describes the provision of information by examination of samples taken from the human or animal body. Most diagnostic SaMD will meet that limb. Products meeting (i) to (iv) instead, such as therapy planning software or software connected with life support, carry a higher risk level, because those purposes describe intervention rather than information. Classification must be argued back to the six purposes one by one. Treating software as a single category and assigning it a class by convention is the error that surfaces at review, when the purpose analysis is requested and cannot be produced. A signed, dated purpose-by-purpose table is the cheapest way to make that argument survive review.

Insight three: paper submission is an additional burden specific to software products. As of 2026 Bangladesh still operates a paper-based submission process, a point covered in the sixteenth article of this series. Software is naturally a digital deliverable, yet in Bangladesh it must be printed, bound into volumes, and accompanied at the packaging stage by a physical carrier and a label. For a product team accustomed to releasing builds, this is the workload most often underestimated. The volume of paper does not shrink because the product is intangible; if anything it grows, since the documentation has to describe behaviour that a reviewer cannot observe directly and must therefore reconstruct from written evidence.

4. Practical recommendations

  • Separate the product form from the delivery form. Treat the product as a device for classification purposes, and treat the delivery as a paper dossier for submission purposes, including printing, binding, and preparing a physical carrier with a label. Printed volumes should follow the structure of the technical file, so that a reviewer can move between them without a separate index.
  • Where the product is deployed on servers outside Bangladesh, prepare a written explanation of data processing and storage in advance. The item has no explicit basis in the guidelines, so it is recorded as pending verification; preparing it costs little, and being asked for it without a prepared answer costs a cycle.
  • Complete the six-purpose classification with particular attention to (i) to (iv) and to (vi), and record the reasoning for the class assigned.
  • Prepare notified body QMS certification for Class B, C and D before assembling the paper file, since the certificate is referenced inside it and cannot be inserted later without re-binding the volume in which it appears.
  • Budget time for printing, binding and courier logistics as project tasks with named owners, rather than as administrative overhead absorbed at the end. A release schedule that ignores printing is a schedule that slips.

5. Verification checklist

  • □ Software confirmed as falling within the device definition, with no claim of software exemption
  • □ Classification completed against the six intended purposes, with attention to (i) to (iv) and (vi)
  • □ Notified body QMS certification prepared for Class B, C and D
  • □ Digital documents converted into bound volumes meeting paper submission requirements
  • □ Data processing explanation for overseas servers listed as a pending verification preparation item

Regulatory information verified as of 27 October 2026.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support across South Asia, including Bangladesh. Our team prepares dossiers to DGDA expectations, manages submissions and deficiency responses, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail